  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76739",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: None    Foreign: France, Germany",
      "recall_number": "Z-1886-2017",
      "product_description": "Stryker SmartLife Large Aseptic Housing (7126-120-000)",
      "product_quantity": "302",
      "reason_for_recall": "Stryker Instruments issued a notice of correction for an update to cleaning practices for the Stryker SmartLife Large Aseptic Housings following complaints of deterioration to the housing and the top and bottom housing separating at the weld. Using cleaning practices outside of the HIFU could lead to housing separating at the weld.",
      "recall_initiation_date": "20170228",
      "center_classification_date": "20170427",
      "termination_date": "20230123",
      "report_date": "20170503",
      "code_info": "Manufacture Start Date for Affected Lots: 18-Mar-2015  Manufacture End Date for Affected Lots: 31-Jan-2017"
    }
  ]
}