      "status": "Terminated",
      "city": "Elgin",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85204",
      "recalling_firm": "Northgate Technologies, Inc.",
      "address_1": "1591 Scottsdale Ct",
      "address_2": "N/A",
      "postal_code": "60123-9361",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution. No governmental.",
      "recall_number": "Z-1885-2020",
      "product_description": "ACMI 9FR PROBE (E4-9F) Manufactured for Olympus Surgical Technologies, America (OSTA); NTI Catalog Number 72-00198-0 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.",
      "product_quantity": "1238 boxes. 3 units per box.",
      "reason_for_recall": "Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.",
      "recall_initiation_date": "20200310",
      "center_classification_date": "20200504",
      "termination_date": "20210202",
      "report_date": "20200513",
      "code_info": "NTI Catalog Number 72-00198-0.",
      "more_code_info": ""
    }
  ]
}