  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plymouth",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73948",
      "recalling_firm": "Greatbatch Medical",
      "address_1": "2300 Berkshire Ln N",
      "address_2": "N/A",
      "postal_code": "55441-4575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: IN, MI, TX.  OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.",
      "recall_number": "Z-1884-2016",
      "product_description": "Torque Limiting Adapter:  Torque Limiting Attachment, Non Sterile, Distributed by Zimmer.    PRECIMED, Torque Limiting Adaptor, Non Sterile..    Customer Number/Model:  00-2360-080-00   T10232,  T116245   T116245, TLA025-25   TLA025-25,  00-2360-080-05   T10233, MAC21003 T116430, TLA025-30 TLA025-30,  MAC21001   T112972, MAC21001   T13621, TLA025-20   TLA025-20.",
      "product_quantity": "5952",
      "reason_for_recall": "Greatbatch Medical has initiated a global Field Recall  for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not  achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction  For Use (IFU) provided with the device.",
      "recall_initiation_date": "20160426",
      "center_classification_date": "20160601",
      "termination_date": "20180607",
      "report_date": "20160608",
      "code_info": "1000207  1001049  1001048  1000698  1000582  1000582  1001334  1001334  1002457  1002456  1002974  1002640  1003175  1003174  1003223  1003223  1003224  1003225  1003548  1003766  1003908  1004435  1004485  1004349  1004903  1005099  1004347  1005530  1005772  1005900  1009013  1008458  1009195  1009195  1009197  1009014  1009198  1009196  1009698  1009199  1009699  1009200  1009700  1011763  1010602  1012455  1012490  1013172  1013173  1013415  1013416  1013921  1013922  1514611  1614605  1714700  1953758  1953809  1953810  2124866  2073128  2116179  2141896  1953758  2150164  2157328  2209891  2302780  2302777  2302781  2302788  2362691  2407040  2407047  2407040  2407040  2898285  2898285  2898289  2971943  2904251  2919512  2938191  2978582  1004138  1004139  1004413  1004578  1004577  1004580  1004579  1005899  1005817  1009701  2324706  2920624  3092836  3092836  1006939      FC-002  FC-002  1012458  6280900001    8027435  8027436  8027434"
    }
  ]
}