{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Road",
      "reason_for_recall": "During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips.  Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the",
      "address_2": "",
      "product_quantity": "53,139 (each) in total",
      "code_info": "Catalog number: 71619 and lot number: ND14 and NE11.",
      "center_classification_date": "20120627",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) including the states of:  AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and the countries of:  BELGIUM, BRAZIL, Byelorussian SSR, CAMBODIA , Dominican Republic, Honduras, Japan, Malaysia, Mexico, Nicaragua, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, UNITED ARAB EMIRATES (UAE), and Vietnam.",
      "state": "MI",
      "product_description": "X-COATED TRANSPLANT PACK    Product Usage:  The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.",
      "report_date": "20120704",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-1882-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61731",
      "termination_date": "20130424",
      "more_code_info": "",
      "recall_initiation_date": "20120509",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}