  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lewisville",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73978",
      "recalling_firm": "Orthofix, Inc",
      "address_1": "3451 Plano Pkwy",
      "address_2": "N/A",
      "postal_code": "75056-9453",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "KS, NV, TX, CA, ID, FL, MA, MI, NJ, AZ, NH, SC, UT, WA",
      "recall_number": "Z-1881-2016",
      "product_description": "Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct.  Rx Only Provided non Sterile.",
      "product_quantity": "36 units",
      "reason_for_recall": "There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.",
      "recall_initiation_date": "20160427",
      "center_classification_date": "20160601",
      "termination_date": "20161123",
      "report_date": "20160608",
      "code_info": "All lots"
    }
  ]
}