  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74164",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AK, AZ, CA, CO, FL, GA, IL, IN, KY, MI, MN, MO, MT, NV, NH, NJ, NY, NC, OK, OR, PA, SD, TX, VA, WA, WI, and the countries of: Alger, Antiqua & Barbuda, Argentina, Australia, Austria, Belarus, Botswana, Brazil, Bulgaria, Canada, China, Columbia, Croatia, Cyprus, Czech Republic, Ecuador Egypt, Estonia, France, Germany, Greece, Hong Kong, India, Indonesia, Iran,  Israel, Italy, Japan, Lithuania, Malaysia, Malta, Mexico, Morocco, Nabibia, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Portugal, Russia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka,Sweden, Switzerland, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, Zimbabwe",
      "recall_number": "Z-1879-2016",
      "product_description": "Monaco RTP System    Product Usage:  The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.",
      "product_quantity": "1401 units",
      "reason_for_recall": "When DICOM exporting a 3D Monaco plan and the \"Composite Field Sequencing\" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined.",
      "recall_initiation_date": "20160518",
      "center_classification_date": "20160602",
      "termination_date": "20210915",
      "report_date": "20160608",
      "code_info": "Monaco versions 5.10.00 and higher"
    }
  ]
}