  "results": [
    {
      "status": "Terminated",
      "city": "Seattle",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74004",
      "recalling_firm": "Natus Medical Incorporated",
      "address_1": "5900 1st Ave S",
      "address_2": "N/A",
      "postal_code": "98108-3248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "distributed in the US including AZ, CA, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WA DC,and WI and in Canada and Singapore.",
      "recall_number": "Z-1878-2016",
      "product_description": "Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44.  Filter Catalogue Part Number: 54343, 54344, and 100105-01.",
      "product_quantity": "21 filters",
      "reason_for_recall": "The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.",
      "recall_initiation_date": "20160422",
      "center_classification_date": "20160601",
      "termination_date": "20161216",
      "report_date": "20160608",
      "code_info": "Filter Serial Numbers: Filter SN: 2522893, 2522897, 2522898, 2522899, 2522900, 2522903, 2522914, 2522915, 2522916, 2522917, 2522918, 2522919, 2522920, 2522921, 2522922, 2522923, 2522924, 2522925, 2522926, 2522930, and 2522931."
    }
  ]
}