  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82937",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide)  including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.",
      "recall_number": "Z-1875-2019",
      "product_description": "ECG Trunk Cables:   REF/Catalog Number\tDescription  2106305-001\tECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft.  2106305-002\tECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft.  2106305-003\tECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft.  2106305-004\tECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft.  2106306-001\tECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft.  2106306-002\tECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft.  2106306-003\tECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft.  2106306-004\tECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft.  2106307-001\tECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft.  2106307-002\tECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft.  2106307-003\tECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft.  2106307-004\tECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft.  2106308-001\tECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft.  2106308-002\tECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft.  2106308-003\tECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft.  2106308-004\tECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft.  2106310-001\tECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft.  2106310-002\tECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft.  2106311-001\tECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft.  2106311-002\tECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft.    The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.",
      "product_quantity": "58,045 units (31,108 US; 26937 OUS) in total",
      "reason_for_recall": "ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.",
      "recall_initiation_date": "20190510",
      "center_classification_date": "20190625",
      "termination_date": "20240524",
      "report_date": "20190703",
      "code_info": "Product does not have lot or serial numbers    REF/Catalog Number GTIN/UDI  2106305-001 00840682137775  2106305-002 00840682137898  2106305-003 none  2106305-004 none  2106306-001 00840682137843  2106306-002 00840682137737  2106306-003 none  2106306-004 none  2106307-001 00840682137836  2106307-002 00840682137751  2106307-003 none  2106307-004 none  2106308-001 00840682137713  2106308-002 00840682137812  2106308-003 none  2106308-004 none  2106310-001 00840682137935  2106310-002 none  2106311-001 00840682137676  2106311-002 none"
    }
  ]
}