  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bonaduz",
      "state": "",
      "country": "Switzerland",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82328",
      "recalling_firm": "Hamilton Medical AG",
      "address_1": "Via Crusch 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT,  and RI.  And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HU, IT, MT, NL. NO, PL.PT, RO, SE, SI, SK, TR, HR. AGO, ARE, ARM, AUS, AZE, BGD, BHR , BLR, BOL, BRA, CAN, CHL, CHN, COL, CRI, DOM, DZA, ECU, EGY, EST, HKG, IDN, IND, IRQ, ISR, JOR, JPN, KAZ, KOR, KWT, LBN, LBY, LKA, LVA, MAR, MDA, MDV, MEX, MNG, MYS, NPL, NZL, OMN, PAK, PAN, PER, PHL, PRI, PSE, QAT, RUS, SAU, SGP, SUR, SYR, THA, TUN, TWN, UKR, URY, USA, VEN, VNM, YEM,  and ZAF.",
      "recall_number": "Z-1874-2019",
      "product_description": "HAMILTON-G5, with software versions less than or equal to 2.60    The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.",
      "product_quantity": "4338 units",
      "reason_for_recall": "New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.",
      "recall_initiation_date": "20190325",
      "center_classification_date": "20190711",
      "termination_date": "20240603",
      "report_date": "20190717",
      "code_info": "Serial numbers 16192 and lower",
      "more_code_info": ""
    }
  ]
}