  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Norcross",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76875",
      "recalling_firm": "CP Medical Inc",
      "address_1": "1775 Corporate Dr Ste 150",
      "address_2": "N/A",
      "postal_code": "30093-2960",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "AL, AZ, AR, CA, CO, CT, FL, ID, IL, IN, KY, LA, MT, NE, NJ, NM, NY, OK, OR, SC, TX, WA and South Africa",
      "recall_number": "Z-1872-2017",
      "product_description": "PGCL [poly (glycolide-co-caprolactone)] Absorbable Suture, Manufacturer CP Medical, Portland OR, Sterile, Product Codes L923, L459, L493, L762, L463, L922, L495, L463, L497, L494, L493",
      "product_quantity": "504 boxes/12 sutures - 6,048 individual sutures",
      "reason_for_recall": "The tensile strength minimum as directed by USP <881> was not met over the length of the product's shelf life.",
      "recall_initiation_date": "20170202",
      "center_classification_date": "20170420",
      "report_date": "20170426",
      "code_info": "Lot Numbers - 150323-04, 150324-04, 150520-10, 150615-01, 150616-08, 150709-01, 151229-64, 160414-69, 160510-59, 160617-52, 160824-50"
    }
  ]
}