  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68405",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-1872-2014",
      "product_description": "Shaver Handpiece Sterilization Tray. Model number 272-700-000.  For sterilization of Stryker Endoscopy arthroscopic shaver.",
      "product_quantity": "1992 trays",
      "reason_for_recall": "STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.",
      "recall_initiation_date": "20140519",
      "center_classification_date": "20140625",
      "termination_date": "20150507",
      "report_date": "20140702",
      "code_info": "Part Number 272-700-000:  Lot number: 20476.1  23541  23837  26403  26404  26405  26676  27009  27518  28341  30248  34355  35932  36756  39705  05J034834  06con05774  06CON06360  06CON06846  06con06869  06con06960  06con08587  25067-1-12  26882-1-1  26882-1-11  26882-1-12  26882-1-13  26882-1-14  35200-1-1  36610-1-1  38820-1-1  39078-1-1  40504-2-1  40675-1-1  41231-1-1  41827-1-1  42085-1-1  42375-1-1  42527-1-1  42789-1-1  43103-1-1  43589-1-1  43716-1-1  43716-1-3  43716-1-4  44568-1-10  44568-1-11  44568-1-2  44568-1-3  44568-1-4  44568-1-6  44568-1-7  44568-1-8  44568-1-9  46227-1-1  47731-1-1  48595-1-1  48596-1-1  48597-1-1  48598-1-1  48599-1-1  48600-1-1  50101-1-1  50628-1-1  50828-1-1  51237-1-1  52439-1-1  55035-1-1  55422-1-1  55462-1-1  56456-1-1  R471611  SS15329."
    }
  ]
}