  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Beverly",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64985",
      "recalling_firm": "SpineFrontier, Inc.",
      "address_1": "500 Cummings Ctr Ste 3500",
      "address_2": "N/A",
      "postal_code": "01915-6516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution including Puerto Rico and the states of CO, KS, MA and TX.",
      "recall_number": "Z-1871-2013",
      "product_description": "Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A    FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.",
      "product_quantity": "13 drill guides",
      "reason_for_recall": "Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending.  As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.",
      "recall_initiation_date": "20121109",
      "center_classification_date": "20130802",
      "termination_date": "20150605",
      "report_date": "20130814",
      "code_info": "2520301, 2520307"
    }
  ]
}