{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Queensbury",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98463",
      "recalling_firm": "Angiodynamics, Inc.",
      "address_1": "603 Queensbury Ave",
      "address_2": "N/A",
      "postal_code": "12804-7619",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.",
      "recall_number": "Z-1869-2026",
      "product_description": "Soft-Vu Angiographic Catheter, Straight Art, Non-Braided;  SOFT-VU STR ART 4F X 65CM 035 NB 0SH;  Catalog No.: 10731401;  Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch);  Box Quantity: 5 pouches;",
      "product_quantity": "220 units (44 boxes)",
      "reason_for_recall": "The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.",
      "recall_initiation_date": "20260303",
      "center_classification_date": "20260417",
      "report_date": "20260429",
      "code_info": "SOFT-VU STR ART 4F X 65CM 035 NB 0SH; Catalog No.: 10731401; Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch); UDI-DI: 25051684009475(Box), 15051684009478(Pouch); Lot No.: A2825042;",
      "more_code_info": ""
    }
  ]
}