  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Atlanta",
      "address_1": "400 Perimeter Center Ter Ne",
      "reason_for_recall": "MOSAIQ does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order.",
      "address_2": "Ste 50",
      "product_quantity": "101",
      "code_info": "Released builds 2.50.04D5 and higher",
      "center_classification_date": "20140625",
      "distribution_pattern": "Worldwide Distribution -  USA (nationwide) including Puerto Rico and the states of  AL, AK, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, IA, KY, LA, ME, MD, MA, MI, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TX, VT, VA and WA,. and the countries of Australia, Canada, New Zealand and United Kingdom.",
      "state": "GA",
      "product_description": "MOSAIQ    MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used where ever radiotherapy and/or chemotherapy are prescribed.",
      "report_date": "20140702",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Elekta, Inc.",
      "recall_number": "Z-1869-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68386",
      "termination_date": "20210716",
      "recall_initiation_date": "20140529",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}