  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plainsboro",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68330",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "311 Enterprise Dr",
      "address_2": "N/A",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1868-2014",
      "product_description": "Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number:      02-6000-050 , Lot Number: 132160.        Product Usage:  Accell Evo3c Demineralized Bone Matrix Putty is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.",
      "product_quantity": "61",
      "reason_for_recall": "This lot of Accell Evo3C Demineralized Bone Matrix, 5cc was incorrectly packaged in boxes that describe the product as Accell Evo3 Demineralized Bone Matrix.  Evo3C is contraindicated for treatment of vertebral compression and not indicated for use in the posterolateral spine.",
      "recall_initiation_date": "20140519",
      "center_classification_date": "20140624",
      "termination_date": "20140905",
      "report_date": "20140702",
      "code_info": "Model Number:      02-6000-050 , Lot Number: 132160."
    }
  ]
}