{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bozeman",
      "state": "MT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65544",
      "recalling_firm": "Mesa Laboratories, Inc., Bozeman Manufacturing Facility",
      "address_1": "10 Evergreen Dr Ste E",
      "address_2": "N/A",
      "postal_code": "59715-2457",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Australia, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Egypt, Greece, India,Israel, Japan, Jordan, Malaysia, Mexico, Milanese, Netherlands, Peru, Portugal, Pozan, Romania, Spain, Sweden, Taiwan, Thailand, United Kingdom, and Uruguay.",
      "recall_number": "Z-1868-2013",
      "product_description": "Mesa Smart-Read EZTEST BIOLOGICAL INDICATOR MONITORING SYSTEM.    Intended for use in determining the efficacy of steam sterilization processes.",
      "product_quantity": "2447/100 count boxes",
      "reason_for_recall": "Self-contained steam biological indicator, are recalled because Mesa recently determined that the read out time is not in specification with the label claims.",
      "recall_initiation_date": "20130604",
      "center_classification_date": "20130801",
      "termination_date": "20131122",
      "report_date": "20130807",
      "code_info": "Lot No SR-444 Manufacture Date 2013 January 21 Expiration Date 2015 January 21;    Lot No SR-445 Manufacture Date 2013 February 15 Expiration Date 2015 February 15.    Lot No SR-447 Manufacture Date 2013 March 18 Expiration Date 2015 March 18.",
      "more_code_info": ""
    }
  ]
}