      "country": "United States",
      "city": "Orlando",
      "address_1": "12151 Research Pkwy Ste 200",
      "reason_for_recall": "The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate.  Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.",
      "address_2": "",
      "product_quantity": "4375 units",
      "code_info": "Model Number: 453564241901  Serial Numbers/UDI (if available):  US1203010330  US1203010371  US1203010373  US1203010382  US1203010400  US1203010413  US1203010421  US1311000007  US1311000009/ (01)00884838083523(21)US1239000009  US1311000017/ (01)00884838083523(21)US1331000034  US1239000009  US1331000034  US1231000061  US1108010053  US1202010217  US1234000060  US1337000038  US1337000040  US1337000043  US1343000024  US1334000021  US1234000066  US1202010242/ (01)00884838083523(21)US1202010242  US1244000002/ (01)00884838083523(21)US1244000002  US1251000087/ (01)00884838083523(21)US1251000087  US1202010168  US1203010284  US1244000061  US1251000005  US1203010302/ (01)00884838083523(21)US1203010302  US1244000016/ (01)00884838083523(21)US1203010340  US1231000038/ (01)00884838083523(21)US1231000038  US1234000024/ (01)00884838083523(21)US1234000024  US1244000016/ (01)00884838083523(21)US1244000016  US1244000040/ (01)00884838083523(21)US1244000040  US1244000066/ (01)00884838083523(21)US1244000066  US1251000013/ (01)00884838083523(21)US1251000013  US1311000025/ (01)00884838083523(21)US1311000025  US1319000059/ (01)00884838083523(21)US1319000059  US1337000032/ (01)00884838083523(21)US1337000032  US1231000048/ (01)00884838084926(21)US1231000048  US1251000031/ (01)00884838084926(21)US1251000031  US1304000040/ (01)00884838084926(21)US1304000040  US1334000041/ (01)00884838084926(21)US1334000041  US1107010016  US1107010019  US1107010022  US1107010036  US1108010062  US1109010104  US1202010153  US1202010173  US1202010233  US1203010261  US1203010295  US1203010307  US1203010320  US1203010345  US1203010377  US1231000022  US1234000013  US1234000049  US1239000024  US1239000045  US1244000032  US1244000055  US1248000023  US1251000030  US1251000033  US1251000052  US1251000077  US1304000020  US1311000069  US1319000009  US1319000016  US1319000030  US1319000069  US1324000008  US1324000014  US1324000039  US1328000002  US1331000040  US1337000055  US1343000001  US1343000009  US1343000027  US1343000059  US1343000073  US1239000002  US1352000047  US1324000011  US1319000038  US1239000019  US1107010007  US1201010117  US1201010119  US1203010401  US1231000026  US1231000049  US1231000059  US1234000051  US1325000009  US1234000086  US1238000005  US1311000011  US1311000062  US1203010378  US1248000041/ (01)00884838083523(21)US1248000041  US1239000011  US1202010208/ 01)00884838083523(21)US1202010208  US1244000054/ (01)00884838083523(21)US1244000054  US1324000049/ (01)00884838083523(21)US1324000049  US1234000068  US1239000008  US1319000091  US1325000017  US1107010011  US1231000058  US1337000024  US1325000007/ (01)00884838083523(21)US1325000007  US1328000014  US1331000067  US1201010129  US1224000051  US1231000008  US1231000016  US1231000028  US1319000024  US1337000071  US1251000053  US1231000046  US1311000002/ (01)00884838083523(21)US1311000002  US1311000054/ (01)00884838083523(21)US1311000054  US1109010076  US1201010113  US1201010116  US1201010144  US1202010189  US1337000028  US1352000020  US1319000097  US1325000002  US1311000029/ (01)00884838083530(21)US1311000029  US1107010020  US1108010040  US1231000036  US1107010008  US1251000010  US1331000038/ (01)00884838083523(21)US1331000038  US1244000022/ (01)00884838083523(21)US1244000022  US1311000058/ (01)00884838083523(21)US1311000058  US1331000057/ (01)00884838083530(21)US1331000057  US1107010037  US1244000019  US1324000032  US1331000028  US1334000032  US1319000047  US1324000029  US1325000020  US1248000019  US1234000048  US1244000042  US1244000053  US1248000032  US1203010291  US1337000019  US1319000033/ (01)00884838083523(21)US1319000033  US1244000036/ (01)00884838084926(21)US1244000036  US1109010092  US1201010143  US1203010386  US1352000003  US1352000012  US1407000022  US1319000060  US1251000050  US1319000001  US1224000044  US1109010098  US1107010015  US1107010023  US1108010041  US1109010068  US1407000002  US1407000029  US1407000031  US1304000027  US1231000067  US1231000073  US1304000034  US1324000030/ (01)00884838083523(21)US1324000030  US1224000012  US1337000050  US1337000066  US1202010149  US1202010176  US1203010348  US1231000034  US1109010081  US1203010265  US1239000006  US1234000021  US1319000015  US1203010316  US1231000010  US1201010132  US1107010012  US1107010017  US1107010021  US1108010059  US1201010115  US1234000083  US1319000052  US1337000018  US1337000048  US1304000011  US1319000029  US1331000004  US1251000038/ (01)00884838083523(21)US1251000038  US1202010196  US1202010197  US1203010354  US1248000008  US1408000013  US1203010357  US1248000018  US1328000012  US1331000033  US1224000002  US1319000094  US1331000016  US1343000056  US1343000020  US1109010107/ (01)00884838083523(21)US1109010107  US1324000018  US1343000060  US1201010138  US1251000073  US1201010123  US1203010351  US1109010086  US1248000056  US1304000002  US1331000052  US1343000012  US1231000042  US1107010027  US1203010391  US1231000062  US1239000026  US1311000018  US1311000026  US1311000044  US1248000017  US1248000066  US1319000020  US1331000064  US1244000018  US1325000022/ (01)00884838083523(21)US1325000022  US1324000022/ (01)00884838084926(21)US1324000022  US1244000067  US1248000006  US1248000013  US1248000020  US1248000049  US1248000051  US1248000062  US1319000005  US1202010200  US1202010177  US1202010191  US1324000033  US1234000087/ (01)00884838083523(21)US1234000087  US1203010264/ (01)00884838083530(21)US1203010264  US1203010376  US1203010388  US1231000047  US1304000026  US1244000046  US1319000026  US1203010389  US1244000060  US1248000060  US1337000020  US1343000065  US1407000054  US1331000018  US1109010066  US1109010077  US1109010111  US1109010088  US1107010001  US1203010266  US1203010321  US1203010390  US1319000098  US1234000072/ (01)00884838083523(21)US1234000072  US1311000021/ (01)00884838083523(21)US1311000021  US1202010205  US1202010210  US1203010329  US1239000025  US1319000004  US1319000084  US1352000009  US1352000063  US1407000055  US1203010288/ (01)00884838083523(21)US1203010288  US1244000028/ (01)00884838083523(21)US1244000028  US1324000012  US1239000015  US1224000025  US1324000006  US1324000051  US1343000050  US1231000030  US1251000047  US1331000017  US1337000030  US1337000036  US1337000051  US1337000054  US1337000062  US1337000063  US1337000064  US1109010094  US1224000053  US1231000044  US1248000055  US1319000054  US1325000016  US1328000005  US1224000043  US1231000086  US1311000005  US1324000013  US1337000059  US1343000071  US1343000077  US1343000091  US1239000050  US1224000001  US1248000054  US1304000013  US1248000043  US1202010201  US1251000017/ (01)00884838083523(21)US1251000017  US1109010103  US1239000032  US1319000011  US1319000027  US1334000002  US1334000027  US1319000086  US1202010203  US1231000040  US1331000009  US1203010253  US1203010258  US1203010303  US1203010344  US1203010368  US1108010050  US1108010063  US1311000030  US1311000008/ (01)00884838083523(21)US1311000008  US1107010024  US1234000059  US1343000066  US1202010241  US1210000005  US1224000011  US1224000019  US1224000054  US1407000006  US1407000047  US1407000050  US1407000051  US1408000005  US1408000021  US1408000022  US1408000027  US1408000028  US1408000030  US1203010412  US1319000032  US1311000034/ (01)00884838083523(21)US1311000034  US1224000040/ (01)00884838083523(21)US1224000040  US1231000003  US1231000005  US1231000006  US1234000028  US1231000053  US1202010158  US1202010159  US1202010162  US1202010163  US1331000054/ (01)00884838084926(21)US1331000054  US1337000014/ (01)00884838084926(21)US1337000014  US1324000028  US1248000047  US1251000029  US1407000039  US1319000055  US1234000030  US1109010071/ (01)00884838083523(21)US1109010071  US1203010272/ (01)00884838083523(21)US1203010272  US1203010296/ (01)00884838083523(21)US1203010296  US1203010394/ (01)00884838083523(21)US1203010394  US1224000008/ (01)00884838083523(21)US1224000008  US1224000026/ (01)00884838083523(21)US1224000026  US1231000009/ (01)00884838083523(21)US1231000009  US1234000002/ (01)00884838083523(21)US1234000002  US1234000007/ (01)00884838083523(21)US1234000007  US1234000082/ (01)00884838083523(21)US1234000082  US1244000052/ (01)00884838083523(21)US1244000052  US1311000051/ (01)00884838083523(21)US1311000051  US1311000071/ (01)00884838083523(21)US1311000071  US1325000024/ (01)00884838083523(21)US1325000024  US1337000006/ (01)00884838083523(21)US1337000006  US1337000073/ (01)00884838083523(21)US1337000073  US1231000015/ (01)00884838083530(21)US1231000015  US1108010046  US1109010096  US1201010131  US1202010170  US1202010188  US1203010252  US1203010300  US1203010331  US1203010370  US1203010403  US1203010405  US1224000022  US1231000024  US1231000079  US1234000043  US1234000052  US1234000063  US1234000073  US1234000090  US1239000001  US1239000010  US1239000037  US1239000041  US1239000043  US1244000015  US1244000020  US1244000037  US1248000022  US1248000033  US1248000040  US1248000061  US1251000007  US1251000035  US1304000033  US1304000044  US1311000020  US1311000022  US1311000032  US1311000053  US1319000021  US1319000040  US1319000063  US1319000089  US1325000008  US1325000015  US1328000001  US1343000069  US1244000035/ (01)00884838083523(21)US1244000035  US1319000013/ (01)00884838083523(21)US1319000013  US1231000070  US1202010167  US1202010182  US1202010193  US1343000004  US1202010225/ (01)00884838083523(21)US1202010225  US1319000023  US1325000004  US1311000067  US1248000048/ (01)00884838083523(21)US1248000048  US1202010183  US1248000053  US1251000037  US1334000024  US1337000060  US1324000026-D  US1202010207  US1244000049  US1201010114  US1201010118  US1319000057  US1201010125  US1203010380  US1251000028  US1324000007  US1248000027  US1203010410  US1244000001  US1244000057  US1319000041  US1337000065  US1352000029  US1352000053  US1352000058  US1311000057  US1234000089/ (01)00884838084926(21)US1234000089  US1203010356  US1201010127  US1202010211  US1203010286  US1203010310  US1203010335  US1244000004  US1244000012  US1311000039  US1331000037  US1224000015/ (01)00884838083523(21)US1224000015  US1109010067  US1109010110/ (01)00884838083523(21)US1109010110  US1248000045/ (01)00884838083523(21)US1248000045  US1325000019  US1202010157/ (01)00884838083523(21)US1202010157  US1319000045/ (01)00884838083523(21)US1319000045  US1231000055  US1239000035  US1311000024  US1319000007  US1324000003  US1324000020  US1201010121  US1203010324  US1224000038  US1234000071  US1248000016  US1203010267  US1244000064  US1311000010  US1319000066  US1251000054/ (01)00884838083523(21)US1251000054  US1352000002  US1108010055  US1201010133  US1201010137  US1202010186  US1203010290  US1203010337  US1337000009  US1324000034  US1324000038/ (01)00884838083523(21)US1324000038  US1203010385  US1224000009  US1231000014  US1234000078  US1324000010  US1324000036  US1324000035/ (01)00884838083523(21)US1324000035  US1202010238  US1248000064  US1311000056  US1412000008  US1234000041  US1107010025  US1107010077  US1334000007  US1352000011  US1352000027  US1203010417/ (01)00884838083523(21)US1203010417  US1203010404  US1352000050  US1202010209/ (01)00884838083523(21)US1202010209  US1337000049/ (01)00884838083523(21)US1337000049  US1248000078  US1251000002  US1251000078  US1319000003  US1319000028  US1325000026  US1331000061  US1107010033  US1251000055  US1337000011  US1203010342  US1108010049/ (01)00884838083523(21)US1108010049  US1234000046/ (01)00884838083523(21)US1234000046  US1319000006/ (01)00884838083523(21)US1319000006  US1319000035  US1343000058  US1311000050  US1238000009  US1331000010  US1234000039  US1234000047  US1234000061  US1239000016  US1108010042  US1202010160  US1202010214  US1231000025  US1248000009  US1203010281  US1203010313  US1203010350  US1203010355  US1324000043/ (01)00884838083523(21)US1324000043  US1352000055  US1202010213  US1202010226  US1202010231  US1234000011  US1304000025  US1231000037/ (01)00884838083523(21)US1231000037  US1203010392/ (01)00884838083523(21)US1203010392  US1244000021/ (01)00884838083523(21)US1244000021  US1201010112  US1231000051  US1337000052  US1319000010/ (01)00884838083523(21)US1319000010  US1203010381  US1231000027  US1337000015  US1244000008  US1334000017  US1244000007  US1325000010  US1331000015  US1331000030  US1334000033  US1334000035  US1334000036  US1334000040  US1334000042  US1334000044  US1337000026  US1231000020  US1248000074  US1231000002  US1324000015  US1202010219  US1202010224  US1202010227  US1202010236  US1202010239  US1202010248  US1203010270  US1224000007  US1202010199  US1304000047  US1311000004  US1328000006  US1202010194  US1203010347  US1238000004  US1319000079  US1331000019  US1331000024  US1331000031  US1331000046  US1331000065  US1234000031  US1251000067  US1107010030  US1319000078  US1109010097  US1224000032  US1239000007  US1331000029  US1234000042    US1311000006/ (01)00884838084926(21)US1311000006  US1319000051/ (01)00884838083523(21)US1319000051  US1311000077/ (01)00884838083523(21)US1311000077  US1328000004  US1251000021  US1324000017  US1343000070  US1234000045  US1324000025  US1203010418  US1334000026  US1334000028  US1203010414  US1319000061  US1331000014  US1108010051  US1108010061  US1337000034/ (01)00884838083523(21)US1337000034  US1109010105  US1203010289  US1203010326  US1319000064  US1324000021  US1311000080  US1231000063  US1109010083/ (01)00884838083523(21)US1109010083  US1107010010/ (01)00884838083523(21)US1107010010  US1244000005  US1109010065  US1248000036  US1109010084  US1202010175/ (01)00884838083523(21)US1202010175  US1201010122  US1248000079  US1325000025  US1201010126/ (01)00884838083523(21)US1201010126  US1234000025  US1343000013/ (01)00884838083523(21)US1343000013  US1244000065  US1334000030/ (01)00884838083523(21)US1334000030  US1108010058/ (01)00884838083523(21)US1108010058  US1201010120  US1202010166  US1251000044  US1324000026  US1304000007  US1234000056/ (01)00884838083523(21)US1234000056  US1331000060  US1311000013  US1202010155/ (01)00884838083523(21)US1202010155  US1251000020/ (01)00884838083523(21)US1251000020  US1337000003/ (01)00884838083523(21)US1337000003  US1337000022/ (01)00884838083523(21)US1337000022  US1337000042  US1343000078  US1337000012  US1203010285  US1224000033  US1234000069/ (01)00884838083523(21)US1234000069  US1343000021/ (01)00884838083523(21)US1343000021  US1304000005/ (01)00884838083523(21)US1304000005  US1231000029  US123100080  US1234000005  US1319000008  US1319000083  US1337000021  US1319000042/ (01)00884838083523(21)US1319000042  US1325000023  US1239000048  US1319000072  US1234000079  US1251000051  US1251000097  US1319000012  US1203010411  US1203010250  US1203010256  US1203010259  US1203010263  US1203010269  US1203010273  US1203010274  US1203010275  US1203010277  US1203010282  US1203010299  US1203010301  US1203010304  US1203010312  US1203010318  US1203010365  US1203010369  US1203010372  US1203010399  US1203010406  US1203010415  US1203010416  US1203010419  US1203010424  US1210000003  US1210000004  US1224000005  US1224000028  US1224000031  US1231000001  US1231000011  US1231000021  US1231000039  US1231000041  US1231000043  US1231000054  US1231000056  US1231000072  US1231000087  US1231000088  US1231000089  US1234000010  US1234000015  US1234000022  US1234000026  US1234000027  US1234000029  US1234000033  US1234000034  US1234000038  US1234000044  US1234000065  US1234000075  US1234000076  US1234000081  US1234000085  US1239000003  US1244000030  US1244000045  US1244000047  US1244000051  US1244000056  US1244000059  US1244000063  US1248000003  US1248000069  US1248000076  US1248000077  US1251000001  US1251000004  US1251000009  US1251000032  US1251000034  US1251000036  US1251000042  US1251000064  US1251000094  US1304000001  US1304000003  US1304000015  US1304000022  US1304000028  US1304000031  US1304000032  US1304000037  US1304000038  US1311000001  US1311000015  US1311000019  US1311000031  US1311000041  US1311000046  US1311000052  US1311000055  US1311000059  US1311000060  US1311000061  US1311000063  US1311000064  US1311000066  US1311000068  US1311000070  US1311000073  US1311000074  US1311000075  US1311000078  US1311000079  US1319000034  US1319000044  US1319000046  US1319000048  US1319000050  US1319000053  US1319000056  US1319000062  US1319000068  US1319000073  US1319000076  US1319000081  US1319000088  US1319000090  US1319000092  US1319000096  US1319000099  US1319000100  US1328000011  US1331000005  US1331000027  US1331000035  US1331000036  US1331000039  US1331000045  US1331000048  US1331000055  US1331000066  US1334000003  US1334000005  US1334000006  US1334000012  US1334000014  US1334000015  US1334000018  US1334000023  US1334000029  US1334000034  US1334000046  US1334000048  US1337000002  US1337000004  US1337000010  US1337000027  US1337000031  US1337000037  US1337000044  US1337000046  US1337000058  US1337000067  US1343000007  US1343000010  US1343000014  US1343000015  US1343000017  US1343000018  US1343000023  US1343000029  US1343000030  US1343000032  US1343000035  US1343000036  US1343000037  US1343000043  US1343000045  US1343000046  US1343000047  US1343000048  US1343000052  US1343000055  US1343000062  US1343000063  US1343000064  US1343000074  US1343000080  US1203010254  US1203010255  US1203010268  US1203010292  US1203010305  US1203010343  US1203010409  US1231000065  US1231000068  US1231000077  US1234000023  US1234000054  US1234000058  US1234000074  US1251000056  US1311000037  US1311000040  US1319000043  US1319000093  US1328000010  US1331000058  US1334000008  US1334000010  US1337000069  US1343000040  US1343000042  US1343000053  US1343000076  US1202010169  US1202010165  US1202010216  US1203010276  US1203010349  US1109010090  US1319000037  US1244000013/ (01)00884838083530(21)US1244000013  US1224000029  US1109010064  US1331000041/ (01)00884838083523(21)US1331000041  US1202010229  US1202010171  US1109010099  US1201010135  US1239000022  US1107010002  US1319000067  US1203010338  US1248000083  US1331000059/ (01)00884838083523(21)US1331000059  US1244000033/ (01)00884838083523(21)US1244000033  US1244000011  US1337000070/ (01)00884838083523(21)US1337000070  US1203010359  US1319000080  US1319000065  US1304000012  US1234000053  US1234000040  US1337000074  US1203010402  US1251000018  US1202010247  US1319000017  US1248000011/ (01)00884838083523(21)US1248000011  US1224000045  US1311000072  US1248000005  US1107010006/ (01)00884838083523(21)US1107010006  US1244000038  US1203010278/ (01)00884838083530(21)US1203010278  US1311000014  US1311000049  US1319000031/ (01)00884838083523(21)US1319000031  US1319000036  US1251000040  US1202010156  US1251000027    Model Number:  453564483321  Serial Numbers/UDI (if available):  US1518000781/ (01)00884838083523(21)US1518000781  US1550001457/ (01)00884838083523(21)US1550001457  US1550001457/ (01)00884838083523(21)US1550001483  US1407000013   US1611001885   US1611001923   US1613001945   US1613001952   US1529001016/ (01)00884838083523(21)US1529001016  US1352000030/ (01)00884838083523(21)US1352000030  US1417000027   US1408000006/ (01)00884838083523(21)US1408000006  US1502000443/ (01)00884838083523(21)US1502000443  US1505000555/ (01)00884838083523(21)US1505000555  US1603001565/ (01)00884838083523(21)US1603001565  US1608001633/ (01)00884838083523(21)US1608001633  US1107010032   US1109010106   US1202010164   US1202010243   US1231000085   US1234000032   US1234000084   US1239000004   US1239000013   US1239000018   US1239000036   US1239000042   US1239000051   US1239000052   US1244000009   US1244000017   US1244000023   US1244000029   US1244000034   US1244000043   US1248000012   US1248000044   US1248000080   US1248000082   US1251000008   US1251000024   US1251000026   US1251000057   US1251000059   US1251000065   US1251000076   US1251000079   US1251000082   US1251000088   US1304000009   US1319000077   US1319000082   US1324000016   US1324000037   US1324000041   US1324000045   US1325000006   US1325000028   US1328000009   US1328000015   US1334000011   US1334000013   US1334000020   US1334000031   US1337000001   US1337000005   US1337000013   US1337000016   US1337000017   US1337000035   US1337000039   US1337000045   US1337000084   US1343000041   US1343000054   US1343000068   US1343000072   US1343000079   US1408000034   US1419000017   US1431000034   US1433000046   US1517000690   US1529001028   US1529001056   US1638002571/ (01)00884838082113(21)US1638002571  US1109010101/ (01)00884838083523(21)US1109010101  US1224000048/ (01)00884838083523(21)US1224000048  US1231000013/ (01)00884838083523(21)US1231000013  US1231000018/ (01)00884838083523(21)US1231000018  US1239000017/ (01)00884838083523(21)US1239000017  US1244000026/ (01)00884838083523(21)US1244000026  US1248000021/ (01)00884838083523(21)US1248000021  US1319000049/ (01)00884838083523(21)US1319000049  US1337000056/ (01)00884838083523(21)US1337000056  US1352000016/ (01)00884838083523(21)US1352000016  US1352000066/ (01)00884838083523(21)US1352000066  US1407000014/ (01)00884838083523(21)US1407000014  US1417000013/ (01)00884838083523(21)US1417000013  US1417000040/ (01)00884838083523(21)US1417000040  US1419000043/ (01)00884838083523(21)US1419000043  US1419000058/ (01)00884838083523(21)US1419000058  US1424000012/ (01)00884838083523(21)US1424000012  US1433000085/ (01)00884838083523(21)US1433000085  US1438000135/ (01)00884838083523(21)US1438000135  US1441000154/ (01)00884838083523(21)US1441000154  US1443000215/ (01)00884838083523(21)US1443000215  US1446000264/(01)00884838083523(21)US1446000264  US1446000301/ (01)00884838083523(21)US1446000301  US1450000347/ (01)00884838083523(21)US1450000347  US1505000538/ (01)00884838083523(21)US1505000538  US1505000554/ (01)00884838083523(21)US1505000554  US1511000603/ (01)00884838083523(21)US1511000603  US1517000662/ (01)00884838083523(21)US1517000662  US1518000747/ (01)00884838083523(21)US1518000747  US1518000756/ (01)00884838083523(21)US1518000756  US1525000989/ (01)00884838083523(21)US1525000989  US1529001019/ (01)00884838083523(21)US1529001019  US1529001047/ (01)00884838083523(21)US1529001047  US1529001075/ (01)00884838083523(21)US1529001075  US1529001081/ (01)00884838083523(21)US1529001081  US1533001114/ (01)00884838083523(21)US1533001114  US1534001202/ (01)00884838083523(21)US1534001202  US1534001205/ (01)00884838083523(21)US1534001205  US1534001231/ (01)00884838083523(21)US1534001231  US1537001273/ (01)00884838083523(21)US1537001273  US1537001307/ (01)00884838083523(21)US1537001307  US1547001368/ (01)00884838083523(21)US1547001368  US1547001390/ (01)00884838083523(21)US1547001390  US1547001432/ (01)00884838083523(21)US1547001432  US1550001495/ (01)00884838083523(21)US1550001495  US1603001547/ (01)00884838083523(21)US1603001547  US1604001578/ (01)00884838083523(21)US1604001578  US1608001659/ (01)00884838083523(21)US1608001659  US1609001690/ (01)00884838083523(21)US1609001690  US1609001705/ (01)00884838083523(21)US1609001705  US1611001829/ (01)00884838083523(21)US1611001829  US1611001834/ (01)00884838083523(21)US1611001834  US1611001852 /(01)00884838083523(21)US1611001852  US1617002011 /(01)00884838083523(21)US1617002011  US1623002178 /(01)00884838083523(21)US1623002178  US1623002184 /(01)00884838083523(21)US1623002184  US1343000003  US1352000034   US1352000037   US1352000061   US1407000037   US1412000044   US1417000005   US1419000028   US1424000007   US1424000013   US1431000016   US1433000052   US1438000091   US1443000211   US1443000223   US1502000465   US1505000530   US1511000580   US1511000602   US1511000608   US1511000614   US1511000643   US1517000658   US1518000764   US1518000796   US1525000923   US1529001044   US1534001153   US1537001286   US1547001404   US1547001429   US1604001592   US1609001723   US1610001764   US1610001782   US1611001855   US1623002127   US1623002134   US1625002340   US1644002650/ (01)00884838083523(21)US1644002650  US1547001385   US1511000645   US1517000655   US1408000035   US1518000789   US1638002590/ (01)00884838083523(21)US1638002590  US1638002596/ (01)00884838083523(21)US1638002596  US1547001421   US1547001426   US1547001433   US1450000345/ (01)00884838083523(21)US1450000345  US1511000574/ (01)00884838083523(21)US1511000574  US1523000873/ (01)00884838083523(21)US1523000873  US1511000599   US1203010423   US1438000120/ (01)00884838083523(21)US1438000120  US1529001057   US1623002123   US1505000496   US1623002137   US1550001447   US1550001466   US1517000668/ (01)00884838083523(21)US1517000668  US1108010054   US1417000019   US1623002115/ (01)00884838083523(21)US1623002115  US1107010038/ (01)00884838083523(21)US1107010038  US1518000786/ (01)00884838083523(21)US1518000786  US1518000801/ (01)00884838083523(21)US1518000801  US1525000912/ (01)00884838083523(21)US1525000912  US1108010056   US1412000011   US1611001893   US1604001580/ (01)00884838083523(21)US1604001580  US1417000032   US1417000038   US1443000177   US1518000784   US1529001088   US1533001103   US1537001310   US1537001314   US1446000299/ (01)00884838083523(21)US1446000299  US1511000604/ (01)00884838083523(21)US1511000604  US1625002284/ (01)00884838083523(21)US1625002284  US1412000020   US1419000006   US1419000047   US1419000056   US1419000059   US1511000635   US1518000768   US1523000839   US1547001357   US1547001420   US1419000031   US1502000380/ (01)00884838083523(21)US1502000380  US1537001288/ (01)00884838083523(21)US1537001288  US1608001646/ (01)00884838083523(21)US1608001646  US1523000885   US1547001383   US1547001438   US1550001446   US1608001650   US1608001657   US1523000848/ (01)00884838083523(21)US1523000848  US1537001258   US1547001416   US1547001439   US1625002257   US1610001801   US1611001860   US1518000785/ (01)00884838083523(21)US1518000785  US1625002237/ (01)00884838083523(21)US1625002237  US1623002154   US1203010314/ (01)00884838083523(21)US1203010314  US1412000046/ (01)00884838083523(21)US1412000046  US1443000169   US1446000295   US1505000537   US1505000545   US1511000592   US1517000650   US1517000673   US1517000687   US1517000709   US1529001049   US1529001059   US1331000026/ (01)00884838083523(21)US1331000026  US1534001181   US1624002222   US1511000633   US1525000921   US1525000939   US1523000828   US1502000472   US1517000735   US1525000904   US1547001397/ (01)00884838083523(21)US1547001397  US1619002060   US1619002061   US1619002102/ (01)00884838083523(21)US1619002102  US1446000257   US1529001032   US1625002238   US1502000453   US1638002614/ (01)00884838083523(21)US1638002614  US1638002547/ (01)00884838083530(21)US1638002547  US1638002577/ (01)00884838083530(21)US1638002577  US1517000683   US1517000699   US1537001298   US1518000765/ (01)00884838083523(21)US1518000765  US1502000434   US1518000783   US1518000795   US1638002559/ (01)00884838083523(21)US1638002559  US1525000922   US1525000961   US1443000222   US1604001623/ (01)00884838083523(21)US1604001623  US1352000056   US1523000849   US1611001902   US1505000506   US1505000510   US1505000550   US1319000025/ (01)00884838083523(21)US1319000025  US1517000719/ (01)00884838083523(21)US1517000719  US1537001318   US1523000809   US1626002421/ (01)00884838083523(21)US1626002421  US1419000001   US1419000002   US1419000019   US1529000992   US1529001051   US1604001616   US1625002266   US1502000462/ (01)00884838083523(21)US1502000462  US1438000116   US1143000182/ (01)00884838083523(21)US1143000182  US1511000626   US1511000640   US1534001235   US1534001238   US1534001240   US1534001241   US1537001261   US1547001415   US1550001449   US1617002020   US1611001868   US1613001927   US1517000737   US1625002240   US1625002242   US1626002401   US1626002407   US1623002132   US1623002158   US1623002186   US1623002193   US1624002212   US1502000454/ (01)00884838083523(21)US1502000454  US1443000197   US1638002615/ (01)00884838083523(21)US1638002615  US1644002543/ (01)00884838083523(21)US1644002543  US1644002675/ (01)00884838083523(21)US1644002675  US1611001872   US1611001908   US1625002298   US1638002537/ (01)00884838083523(21)US1638002537  US1638002548/ (01)00884838083523(21)US1638002548  US1638002550/ (01)00884838083523(21)US1638002550  US1638002610/ (01)00884838083523(21)US1638002610  US1412000049   US1623002109   US1623002152   US1624002213   US1625002227   US1625002234   US1625002241   US1625002244   US1625002263   US1550001482   US1550001488   US1609001716/ (01)00884838083523(21)US1609001716  US1424000036   US1417000034   US1419000021   US1424000002   US1438000115   US1523000867   US1604001582   US1604001590   US1331000020/ (01)00884838083523(21)US1331000020  US1407000052/ (01)00884838083523(21)US1407000052  US1502000467/ (01)00884838083523(21)US1502000467  US1523000864/ (01)00884838083523(21)US1523000864  US1529001064/ (01)00884838083523(21)US1529001064  US1431000037   US1433000049   US1433000050   US1505000536   US1608001628   US1517000689/ (01)00884838083523(21)US1517000689  US1547001398/ (01)00884838083523(21)US1547001398  US1412000036   US1517000667   US1611001871   US1625002275   US1625002277   US1604001600   US1604001614   US1613001936   US1613001960   US1644002624/ (01)00884838083523(21)US1644002624  US1644002638/ (01)00884838083523(21)US1644002638  US1623002133   US1623002163   US1623002171   US1624002216   US1624002219   US1408000025   US1324000009/ (01)00884838083523(21)US1324000009  US1419000009/ (01)00884838083523(21)US1419000009  US1446000240/ (01)00884838083523(21)US1446000240  US1511000637/ (01)00884838083523(21)US1511000637  US1517000686/ (01)00884838083523(21)US1517000686  US1523000820/ (01)00884838083523(21)US1523000820  US1529001037/ (01)00884838083523(21)US1529001037  US1529001045/ (01)00884838083523(21)US1529001045  US1529001054/ (01)00884838083523(21)US1529001054  US1529001013   US1529001072   US1550001475   US1533001102   US1610001805/ (01)00884838083523(21)US1610001805  US1352000043/ (01)00884838083523(21)US1352000043  US1424000049/ (01)00884838083523(21)US1424000049  US1443000198/ (01)00884838083523(21)US1443000198  US1511000596/ (01)00884838083523(21)US1511000596  US1523000837/ (01)00884838083523(21)US1523000837  US1608001631/ (01)00884838083523(21)US1608001631  US1352000005   US1412000037   US1417000014   US1505000559   US1604001569   US1604001608   US1352000033   US1604001603   US1604001615   US1446000254   US1446000256   US1446000266   US1502000397   US1529001087   US1529001089   US1603001541   US1547001366   US1608001658   US1547001414   US1525000990   US1529000997   US1529001055   US1502000474   US1534001210/ (01)00884838083523(21)US1534001210  US1623002143/ (01)00884838083523(21)US1623002143  US1505000525   US1517000674/ (01)00884838083523(21)US1517000674  US1534001221/ (01)00884838083523(21)US1534001221  US1608001671/ (01)00884838083523(21)US1608001671  US1424000042   US1518000802   US1547001407   US1534001179/ (01)00884838083523(21)US1534001179  US1625002232/ (01)00884838083523(21)US1625002232  US1502000446   US1517000652/ (01)00884838083523(21)US1517000652  US1611001847/ (01)00884838083523(21)US1611001847  US1638002581/ (01)00884838083523(21)US1638002581  US1619002050   US1523000888/ (01)00884838083523(21)US1523000888  US1511000624   US1518000745/ (01)00884838083523(21)US1518000745  US1244000039/ (01)00884838083523(21)US1244000039  US1343000031   US1408000050   US1550001494   US1604001583   US1609001692   US1609001696   US1609001699   US1609001691   US1626002431/ (01)00884838083523(21)US1626002431  US1626002454/ (01)00884838083523(21)US1626002454  US1352000067   US1408000026   US1533001098   US1331000043/ (01)00884838083523(21)US1331000043  US1511000605   US1533001100   US1610001753   US1433000074   US1433000087   US1523000872   US1502000471/ (01)00884838083523(21)US1502000471  US1417000037   US1511000584   US1511000618   US1433000054   US1518000780   US1109010102/ (01)00884838083523(21)US1109010102  US1502000459/ (01)00884838083523(21)US1502000459  US1609001698/ (01)00884838083523(21)US1609001698  US1625002292   US1529001067/ (01)00884838083523(21)US1529001067  US1511000591   US1525000946   US1550001464   US1550001504   US1234000050/ (01)00884838083523(21)US1234000050  US1352000051/ (01)00884838083523(21)US1352000051  US1408000048/ (01)00884838083523(21)US1408000048  US1613001926   US1613001973   US1517000660/ (01)00884838083523(21)US1517000660  US1550001477   US1611001819   US1611001846   US1611001856   US1617001996   US1617001999   US1534001194   US1534001208   US1431000024   US1438000145/ (01)00884838083523(21)US1438000145  US1610001772   US1533001116   US1533001140   US1231000080   US1638002573/ (01)00884838083523(21)US1638002573  US1109010095/ (01)00884838083523(21)US1109010095  US1239000005/ (01)00884838083523(21)US1239000005  US1408000010/ (01)00884838083523(21)US1408000010  US1517000729/ (01)00884838083523(21)US1517000729  US1537001324/ (01)00884838083523(21)US1537001324  US1547001364/ (01)00884838083523(21)US1547001364  US16230",
      "center_classification_date": "20180514",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico., and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Kazakhstan, Kenya, Malaysia, Mexico, Mauritius, Netherlands, New Zealand, Pakistan, Philippines, Poland, Romania, Russian Fed, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom, Utd. Arab Emir, Uzbekistan and Vietnam.",
      "state": "FL",
      "product_description": "PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers:  453564241901, 453564483321, 453564621791, and 989803199561 (international only)  The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers.",
      "report_date": "20180523",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Invivo Corporation",
      "recall_number": "Z-1867-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79611",
      "termination_date": "20210430",
      "more_code_info": "02168/ (01)00884838083523(21)US1623002168  US1537001260   US1537001266   US1537001300   US1547001410   US1433000088/ (01)00884838083523(21)US1433000088  US1604001570   US1604001613   US1525000905   US1623002129   US1511000623/ (01)00884838083523(21)US1511000623  US1533001111   US1533001115   US1533001118   US1533001121   US1533001125   US1533001126   US1533001132   US1533001133   US1533001134   US1534001201   US1534001203   US1534001214   US1534001216   US1412000013/ (01)00884838083523(21)US1412000013  US1231000045   US1523000811   US1443000196/ (01)00884838083523(21)US1443000196  US1529001084/ (01)00884838083523(21)US1529001084  US1604001602/ (01)00884838083523(21)US1604001602  US1408000020   US1443000194   US1518000794   US1533001137/ (01)00884838083523(21)US1533001137  US1518000791   US1518000793   US1518000797   US1518000798   US1525000943   US1525000963   US1525000966   US1626002415/ (01)00884838083523(21)US1626002415  US1626002432/ (01)00884838083523(21)US1626002432  US1626002451/ (01)00884838083523(21)US1626002451  US1626002455/ (01)00884838083523(21)US1626002455  US1626002465/ (01)00884838083523(21)US1626002465  US1502000424/ (01)00884838083523(21)US1502000424  US1502000450/ (01)00884838083523(21)US1502000450  US1502000451/ (01)00884838083523(21)US1502000451  US1533001148/ (01)00884838083523(21)US1533001148  US1523000870   US1603001518   US1603001519   US1603001551   US1603001555   US1603001557   US1604001586   US1604001597   US1604001598   US1604001599   US1604001604   US1604001621   US1613001938   US1623002128   US1623002131   US1408000023   US1424000021/ (01)00884838083523(21)US1424000021  US1505000561/ (01)00884838083523(21)US1505000561  US1424000040   US1443000191   US1617001995   US1617002001   US1619002055   US1619002062   US1619002064   US1619002068   US1446000294/ (01)00884838083523(21)US1446000294  US1443000173   US1619002058   US1619002070   US1619002082   US1619002085   US1438000108/ (01)00884838083523(21)US1438000108  US1408000031   US1608001662   US1609001693   US1609001700   US1609001715   US1604001584   US1419000012   US1419000025   US1529001014   US1624002207   US1534001182   US1534001186   US1537001299   US1644002651/ (01)00884838082113(21)US1644002651  US1502000396/ (01)00884838083523(21)US1502000396  US1626002378/ (01)00884838083530(21)US1626002378  US1626002458/ (01)00884838083530(21)US1626002458  US1604001574/ (01)00884838083523(21)US1604001574  US1623002122/ (01)00884838083523(21)US1623002122  US1319000074   US1424000009   US1608001675   US1613001954   US1613001958   US1613001977   US1331000051/ (01)00884838083523(21)US1331000051  US1419000030   US1443000209   US1511000613   US1603001550   US1604001606   US1625002243   US1537001263   US1623002120   US1619002067   US1623002179   US1623002183   US1505000492   US1505000533   US1505000535   US1625002230   US1623002108   US1625002245   US1625002250   US1433000045   US1433000047   US1547001412   US1238000006/ (01)00884838083523(21)US1238000006  US1547001379   US1547001392   US1417000035/ (01)00884838083523(21)US1417000035  US1525000909/ (01)00884838083523(21)US1525000909  US1525000959/ (01)00884838083523(21)US1525000959  US1251000069   US1517000678   US1431000003/ (01)00884838083523(21)US1431000003  US1613001965   US1419000015   US1547001427   US1625002233   US1534001227   US1431000001   US1433000076   US1610001788   US1617002006   US1617002036   US1419000029/ (01)00884838083523(21)US1419000029  US1443000167   US1611001925   US1505000503   US1518000792   US1638002498/ (01)00884838083523(21)US1638002498  US1638002511/ (01)00884838083523(21)US1638002511  US1511000646   US1517000724   US1517000730   US1517000732   US1239000023/ (01)00884838083523(21)US1239000023  US1446000262   US1450000373   US1525000911   US1525000955   US1525000979   US1431000013   US1424000017   US1502000387   US1502000392   US1502000412   US1502000390   US1603001520   US1603001526   US1603001527   US1603001532   US1603001539   US1610001803   US1611001808   US1511000620/ (01)00884838083523(21)US1511000620  US1638002582/ (01)00884838083523(21)US1638002582  US1644002648/ (01)00884838083523(21)US1644002648  US1644002669/ (01)00884838083523(21)US1644002669  US1533001106   US1613001984   US1617002000   US1617002015   US1623002114   US1623002125   US1623002126   US1623002191   US1624002202   US1624002205   US1529001024/ (01)00884838083523(21)US1529001024  US1438000146/ (01)00884838083523(21)US1438000146  US1533001094   US1534001166   US1534001168   US1534001174   US1534001211   US1352000039/ (01)00884838083523(21)US1352000039  US1547001437   US1604001581   US1608001647   US1608001666   US1617002027   US1424000008/ (01)00884838083523(21)US1424000008  US1638002603/ (01)00884838083523(21)US1638002603  US1638002604/ (01)00884838083523(21)US1638002604  US1417000016   US1517000654   US1517000693   US1525000958   US1525000964   US1547001406   US1623002194   US1624002196   US1624002217   US1624002218   US1624002221   US1610001762/ (01)00884838083523(21)US1610001762  US1626002461/ (01)00884838083523(21)US1626002461  US1638002551/ (01)00884838083523(21)US1638002551  US1525000920   US1525000930   US1525000936   US1625002264   US1625002293   US1424000014/ (01)00884838083523(21)US1424000014  US1505000479/ (01)00884838083523(21)US1505000479  US1533001150/ (01)00884838083523(21)US1533001150  US1604001620   US1608001655   US1609001695   US1609001733   US1412000003/ (01)00884838083523(21)US1412000003  US1610001755/ (01)00884838083523(21)US1610001755  US1534001192   US1617002034/ (01)00884838083523(21)US1617002034  US1547001376   US1547001393   US1319000075   US1529001086   US1534001187   US1546001355   US1638002561/ (01)00884838083523(21)US1638002561  US1638002600/ (01)00884838083523(21)US1638002600  US1644002649/ (01)00884838083530(21)US1644002649  US1644002664/ (01)00884838083530(21)US1644002664  US1352000025   US1407000026   US1407000034   US1407000036   US1407000049   US1534001180   US1424000043   US1433000042   US1433000043   US1617001993   US1617002017   US1617002039   US1617002044   US1619002084   US1417000004/ (01)00884838083523(21)US1417000004  US1408000032   US1525000985   US1438000106   US1438000134   US1443000213   US1505000543   US1438000125/ (01)00884838083523(21)US1438000125  US1617001990/ (01)00884838083523(21)US1617001990  US1343000044   US1547001381   US1550001465   US1550001486   US1511000638/ (01)00884838083523(21)US1511000638  US1603001521/ (01)00884838083523(21)US1603001521  US1441000152   US1518000790   US1450000346/ (01)00884838083523(21)US1450000346  US1613001974   US1617002007   US1617002019   US1617002024   US1617002037   US1617002042   US1623002180   US1625002271   US1625002272   US1450000354   US1450000364   US1417000036/ (01)00884838083523(21)US1417000036  US1511000579   US1517000681   US1511000598   US1517000697   US1517000704   US1625002231   US1625002239   US1625002253   US1625002255   US1625002259   US1547001365   US1603001510   US1603001514   US1625002262   US1446000286/ (01)00884838083523(21)US1446000286  US1617002004   US1617002018   US1617002023   US1617002041   US1609001734/ (01)00884838083523(21)US1609001734  US1638002505/ (01)00884838083530(21)US1638002505  US1638002543/ (01)00884838083530(21)US1638002543  US1638002568/ (01)00884838083530(21)US1638002568  US1638002594/ (01)00884838083530(21)US1638002594  US1638002606/ (01)00884838083530(21)US1638002606  US1547001440   US1547001409   US1619002047   US1619002048   US1619002057   US1619002090   US1623002167   US1623002173   US1623002182   US1234000036/ (01)00884838083523(21)US1234000036  US1638002563/ (01)00884838083523(21)US1638002563  US1529001077   US1537001283   US1537001303   US1537001305   US1537001315   US1251000071/ (01)00884838083523(21)US1251000071  US1109010089   US1224000014/ (01)00884838083523(21)US1224000014  US1431000029/ (01)00884838083523(21)US1431000029  US1441000159   US1443000174   US1537001275   US1537001308   US1550001467   US1550001498   US1529001078   US1534001185   US1547001377   US1547001402   US1502000423   US1610001778/ (01)00884838083523(21)US1610001778  US1518000740   US1424000046   US1424000048   US1603001512   US1603001522   US1417000021   US1505000515   US1505000548   US1443000188   US1547001430   US1550001503/ (01)00884838083523(21)US1550001503  US1609001688/ (01)00884838083523(21)US1609001688  US1611001817/ (01)00884838083523(21)US1611001817  US1248000038   US1407000030   US1603001511   US1517000707   US1550001489   US1603001516   US1324000050/ (01)00884838083523(21)US1324000050  US1424000023/ (01)00884838083523(21)US1424000023  US1534001173/ (01)00884838083523(21)US1534001173  US1547001389/ (01)00884838083523(21)US1547001389  US1611001809/ (01)00884838083523(21)US1611001809  US1419000039   US1424000024   US1604001619   US1525000972   US1525000944   US1443000186   US1511000644   US1644002647/ (01)00884838082113(21)US1644002647  US1638002558/ (01)00884838083523(21)US1638002558  US1608001630   US1608001684   US1609001687   US1613001959   US1625002260   US1603001506   US1603001525   US1529001042/ (01)00884838083523(21)US1529001042  US1603001515   US1626002373   US1626002443   US1626002444   US1623002156   US1624002210   US1525000906/ (01)00884838083523(21)US1525000906  US1608001635/ (01)00884838083523(21)US1608001635  US1417000031   US1450000344   US1505000489   US1505000493   US1419000020/ (01)00884838083523(21)US1419000020  US1431000025/ (01)00884838083523(21)US1431000025  US1518000788   US1525000987   US1523000825/ (01)00884838083523(21)US1523000825  US1408000008   US1412000033   US1446000247   US1511000634   US1517000705   US1525000976   US1109010074   US1224000030   US1239000530   US1234000009   US1234000019   US1343000051/ (01)00884838083523(21)US1343000051  US1352000038/ (01)00884838083523(21)US1352000038  US1352000040/ (01)00884838083523(21)US1352000040  US1417000007/ (01)00884838083523(21)US1417000007  US1424000004/ (01)00884838083523(21)US1424000004  US1424000037/ (01)00884838083523(21)US1424000037  US1443000203/ (01)00884838083523(21)US1443000203  US1446000305/ (01)00884838083523(21)US1446000305  US1450000338/ (01)00884838083523(21)US1450000338  US1502000420/ (01)00884838083523(21)US1502000420  US1518000803/ (01)00884838083523(21)US1518000803  US1518000807/ (01)00884838083523(21)US1518000807  US1523000884/ (01)00884838083523(21)US1523000884  US1529000993/ (01)00884838083523(21)US1529000993  US1547001380/ (01)00884838083523(21)US1547001380  US1550001485/ (01)00884838083523(21)US1550001485  US1611001835/ (01)00884838083523(21)US1611001835  US1626002384/ (01)00884838083523(21)US1626002384  US1234000067   US1251000066   US1331000032   US1337000061   US1352000013   US1408000037   US1412000021   US1417000026   US1417000041   US1424000005   US1424000025   US1424000027   US1424000035   US1431000017   US1433000086   US1438000129   US1443000200   US1502000448   US1502000475   US1511000581   US1518000758   US1523000886   US1523000887   US1525000938   US1525000988   US1534001164   US1534001189   US1547001378   US1547001396   US1604001571   US1604001588   US1604001601   US1608001627   US1608001685   US1609001725   US1623002121   US1624002197   US1624002225   US1441000160   US1443000204   US1443000207   US1203010367   US1517000731   US1517000734   US1523000889   US1523000894   US1523000897   US1525000919   US1525000925   US1525000934   US1417000028   US1419000037   US1419000041   US1419000050   US1443000218/ (01)00884838083523(21)US1443000218  US1443000168   US1201010145   US1529001079/ (01)00884838083523(21)US1529001079  US1547001417/ (01)00884838083523(21)US1547001417  US1502000445   US1511000586/ (01)00884838083523(21)US1511000586  US1433000053   US1518000766/ (01)00884838083523(21)US1518000766  US1417000017   US1443000220   US1523000827   US1617002035   US1626002394   US1518000804/ (01)00884838083523(21)US1518000804  US1604001593   US1604001596   US1604001605   US1412000012/ (01)00884838083523(21)US1412000012  US1626002420/ (01)00884838083523(21)US1626002420  US1626002430/ (01)00884838083523(21)US1626002430  US1438000105/ (01)00884838083523(21)US1438000105  US1502000391/ (01)00884838083523(21)US1502000391  US1610001774/ (01)00884838083523(21)US1610001774  US1502000436   US1502000455   US1502000463   US1502000473   US1505000563   US1511000578   US1511000611   US1623002175   US1624002214   US1324000001/ (01)00884838083523(21)US1324000001  US1611001922/ (01)00884838083523(21)US1611001922  US1638002565/ (01)00884838083523(21)US1638002565  US1352000010/ (01)00884838083523(21)US1352000010  US1613001983/ (01)00884838083523(21)US1613001983  US1417000022   US1424000032   US1417000024   US1604001572   US1609001732/ (01)00884838083523(21)US1609001732  US1611001882/ (01)00884838083523(21)US1611001882  US1412000048   US1529001050   US1623002107   US1517000703   US1523000813   US1450000336/ (01)00884838083523(21)US1450000336  US1534001224/ (01)00884838083523(21)US1534001224  US1638002510/ (01)00884838083523(21)US1638002510  US1534001175   US1534001212   US1550001474   US1511000577/ (01)00884838083523(21)US1511000577  US1638002601/ (01)00884838083523(21)US1638002601  US1638002618/ (01)00884838083523(21)US1638002618  US1610001773   US1611001839   US1624002203   US1412000009   US1433000068   US1446000249   US1523000900   US1525000973   US1525000978   US1443000214/ (01)00884838083523(21)US1443000214  US1517000701   US1517000708   US1407000021   US1443000176   US1619002097   US1619002104   US1623002110   US1623002119   US1624002208   US1331000042/ (01)00884838083523(21)US1331000042  US1234000035   US1505000532   US1505000558   US1517000733   US1525000917   US1533001119   US1533001128   US1638002616/ (01)00884838083523(21)US1638002616  US1644002631/ (01)00884838083523(21)US1644002631  US1431000028   US1502000460   US1610001779   US1525000910/ (01)00884838083523(21)US1525000910  US1239000029   US1511000648   US1523000859   US1525000962   US1529001029   US1529001031   US1529001036   US1604001594   US1644002636/ (01)00884838083530(21)US1644002636  US1644002642/ (01)00884838083530(21)US1644002642  US1603001531   US1603001533   US1408000044/ (01)00884838083523(21)US1408000044  US1412000043/ (01)00884838083523(21)US1412000043  US1626002426/ (01)00884838083523(21)US1626002426  US1638002502/ (01)00884838083523(21)US1638002502  US1417000025/ (01)00884838083523(21)US1417000025  US1617001986   US1419000040/ (01)00884838083523(21)US1419000040  US1517000694   US1626002404/ (01)00884838083523(21)US1626002404  US1626002411/ (01)00884838083523(21)US1626002411  US1626002412/ (01)00884838083523(21)US1626002412  US1523000883   US1518000777/ (01)00884838083523(21)US1518000777  US1604001576   US1611001806   US1611001897   US1613001956   US1319000019/ (01)00884838083523(21)US1319000019  US1424000033/ (01)00884838083523(21)US1424000033  US1518000782/ (01)00884838083523(21)US1518000782  US1529001021/ (01)00884838083523(21)US1529001021  US1518000763   US1525000940   US1525000960   US1525000968   US1529001005   US1529001010   US1533001144   US1534001177   US1623002146   US1638002560/ (01)00884838083530(21)US1638002560  US1638002597/ (01)00884838083530(21)US1638002597  US1638002598/ (01)00884838083530(21)US1638002598  US1638002612/ (01)00884838083530(21)US1638002612  US1505000491   US1505000500   US1505000516   US1505000527   US1505000549   US1529001046   US1517000695   US1517000698   US1511000615/ (01)00884838083523(21)US1511000615  US1626002408/ (01)00884838083523(21)US1626002408  US1626002422/ (01)00884838083523(21)US1626002422  US1626002427/ (01)00884838083523(21)US1626002427  US1352000026/ (01)00884838083523(21)US1352000026  US1424000050/ (01)00884838083523(21)US1424000050  US1505000560/ (01)00884838083523(21)US1505000560  US1523000863/ (01)00884838083523(21)US1523000863  US1352000018   US1613001964   US1617002008   US1617002021   US1617002026   US1610001796   US1610001804   US1417000042   US1450000322   US1608001648   US1608001649   US1608001653   US1608001663   US1608001683   US1626002397/ (01)00884838083523(21)US1626002397  US1610001785   US1613001978   US1617001992   US1617002014   US1617002033   US1617002040   US1625002294   US1417000030   US1438000103   US1438000139   US1441000151   US1438000107   US1325000011/ (01)00884838083523(21)US1325000011  US1412000027/ (01)00884838083523(21)US1412000027  US1626002387   US1626002390   US1626002417   US1626002424   US1623002166/ (01)00884838083523(21)US1623002166  US1525000974   US1446000278   US1511000647   US1517000661   US1547001371   US1529001085/ (01)00884838083523(21)US1529001085  US1431000035   US1502000466   US1505000501   US1517000659   US1547001356   US1547001388   US1419000011/ (01)00884838083523(21)US1419000011  US1547001408   US1408000003/ (01)00884838083523(21)US1408000003  US1511000590   US1511000642   US1625002291   US1412000016   US1511000585   US1517000700   US1610001781   US1610001791   US1610001797   US1610001798   US1611001815   US1611001816   US1613001955   US1525000984/ (01)00884838083523(21)US1525000984  US1419000018   US1625002261   US1625002290   US1251000061   US1533001099   US1533001104   US1533001110   US1533001117   US1533001129   US1533001130   US1534001170   US1537001301   US1613001953   US1324000024/ (01)00884838083523(21)US1324000024  US1433000083   US1511000589   US1251000058   US1517000710   US1525000942   US1609001724   US1610001748   US1610001775   US1419000036/ (01)00884838083523(21)US1419000036  US1450000369/ (01)00884838083523(21)US1450000369  US1626002418/ (01)00884838083523(21)US1626002418  US1626002457/ (01)00884838083523(21)US1626002457  US1626002466/ (01)00884838083523(21)US1626002466  US1417000020/ (01)00884838083523(21)US1417000020  US1623002149/ (01)00884838083523(21)US1623002149  US1617001994   US1617002032   US1625002281   US1224000018   US1518000741/ (01)00884838083523(21)US1518000741  US1611001919   US1613001944   US1613001972   US1613001975   US1610001771/ (01)00884838083523(21)US1610001771  US1625002256/ (01)00884838083523(21)US1625002256  US1502000402   US1502000429   US1502000432   US1311000028   US1626002385/ (01)00884838083523(21)US1626002385  US1626002393/ (01)00884838083523(21)US1626002393  US1626002402/ (01)00884838083523(21)US1626002402  US1626002409/ (01)00884838083523(21)US1626002409  US1626002435/ (01)00884838083523(21)US1626002435  US1107010026   US1450000376   US1502000399   US1441000161   US1525000975   US1529000999   US1529001000   US1529001001   US1529001004   US1433000066   US1417000002/ (01)00884838083523(21)US1417000002  US1638002554/ (01)00884838083523(21)US1638002554  US1638002576/ (01)00884838083523(21)US1638002576  US1638002589/ (01)00884838083523(21)US1638002589  US1638002592/ (01)00884838083523(21)US1638002592  US1533001093   US1638002585/ (01)00884838083523(21)US1638002585  US1638002587/ (01)00884838083523(21)US1638002587  US1644002633/ (01)00884838083523(21)US1644002633  US1441000157   US1419000005   US1505000557   US1446000273   US1446000285   US1446000291   US1644002632/ (01)00884838082113(21)US1644002632  US1626002434   US1626002459   US1419000053   US1608001634   US1609001697   US1609001726   US1609001731   US1505000573/ (01)00884838083523(21)US1505000573  US1537001321/ (01)00884838083523(21)US1537001321  US1525000956   US1547001358   US1547001370   US1547001373   US1547001391   US1547001401   US1547001435   US1617002013   US1617002016   US1638002562/ (01)00884838083523(21)US1638002562  US1638002602/ (01)00884838083523(21)US1638002602  US1534001199   US1611001883   US1502000417/ (01)00884838083523(21)US1502000417  US1408000016   US1446000274   US1502000440   US1352000007   US1412000026   US1412000030   US1523000810   US1547001423   US1623002157   US1623002161   US1624002211   US1625002251   US1625002274   US1625002287   US1239000012   US1625002312   US1625002318   US1625002320   US1626002372   US1608001680/ (01)00884838083523(21)US1608001680  US1446000259   US1450000361   US1638002488/ (01)00884838083523(21)US1638002488  US1638002497/ (01)00884838083523(21)US1638002497  US1611001875   US1613001937   US1443000217/ (01)00884838083523(21)US1443000217  US1611001870   US1611001877   US1611001901   US1613001929   US1613001948   US1438000136   US1623002124   US1231000076/ (01)00884838083523(21)US1231000076  US1534001183/ (01)00884838083523(21)US1534001183  US1502000464   US1523000817   US1537001325   US1608001639   US1417000008/ (01)00884838083523(21)US1417000008  US1610001760/ (01)00884838083523(21)US1610001760  US1625002330/ (01)00884838083523(21)US1625002330  US1613001966/ (01)00884838083523(21)US1613001966  US1419000038/ (01)00884838083523(21)US1419000038  US1505000562/ (01)00884838083523(21)US1505000562  US1608001670   US1611001910   US1611001918   US1443000190   US1523000896/ (01)00884838083523(21)US1523000896  US1529001012   US1518000778   US1424000041   US1443000199   US1446000255/ (01)00884838083523(21)US1446000255  US1446000269/ (01)00884838083523(21)US1446000269  US1537001330/ (01)00884838083523(21)US1537001330  US1412000042   US1517000714   US1517000727   US1638002492/ (01)00884838083523(21)US1638002492  US1638002506/ (01)00884838083523(21)US1638002506  US1638002521/ (01)00884838083523(21)US1638002521  US1638002525/ (01)00884838083523(21)US1638002525  US1441000156/ (01)00884838083523(21)US1441000156  US1533001096/ (01)00884838083523(21)US1533001096  US1603001508/ (01)00884838083523(21)US1603001508  US1603001529/ (01)00884838083523(21)US1603001529  US1518000755   US1438000150/ (01)00884838083523(21)US1438000150  US1534001160   US1534001218   US1319000039   US1529001006   US1502000442/ (01)00884838083523(21)US1502000442  US1505000570   US1505000571   US1505000572   US1511000587   US1511000588   US1511000593   US1604001575   US1419000032   US1625002267   US1433000062   US1248000029/ (01)00884838083523(21)US1248000029  US1443000206/ (01)00884838083523(21)US1443000206  US1604001595/ (01)00884838083523(21)US1604001595  US1412000005   US1412000025   US1417000015   US1417000018   US1450000342   US1518000771   US1523000812   US1603001538   US1304000042/ (01)00884838083523(21)US1304000042  US1608001669/ (01)00884838083523(21)US1608001669  US1352000068   US1431000023   US1611001886   US1518000770   US1523000869   US1604001617   US1604001622   US1502000422/ (01)00884838083523(21)US1502000422  US1533001146   US1533001151   US1626002399/ (01)00884838083523(21)US1626002399  US1626002383   US1626002391   US1626002396   US1626002398   US1626002410   US1626002439   US1603001528   US1603001530   US1626002381/ (01)00884838083523(21)US1626002381  US1626002419/ (01)00884838083523(21)US1626002419  US1408000024   US1617002043   US1619002054   US1248000039/ (01)00884838083523(21)US1248000039  US1533001095   US1537001281   US1537001296   US1537001323   US1505000551   US1505000568   US1505000481   US1505000486   US1505000519   US1525000916   US1626002376   US1626002433   US1550001442   US1550001443   US1550001448   US1550001450   US1550001455   US1550001460   US1617002009   US1617002029   US1638002528/ (01)00884838083523(21)US1638002528  US1525000908/ (01)00884838083523(21)US1525000908  US1502000384   US1502000395   US1304000036   US1623002153   US1625002297   US1525000970/ (01)00884838083523(21)US1525000970  US1608001632   US1608001664   US1625002249/ (01)00884838083523(21)US1625002249  US1319000070   US1443000178     US1617001998   US1617002022   US1617002030   US1431000038   US1433000070   US1433000072   US1438000144/ (01)00884838083523(21)US1438000144  US1625002282   US1625002345   US1626002386   US1626002423   US1626002447   US1611001909   US1613001930   US1613001962   US1625002247   US1424000047   US1424000052   US1424000054   US1446000267   US1611001845   US1611001821   US1611001838   US1517000711   US1525000929   US1523000901   US1525000914   US1525000949   US1623002160/ (01)00884838083523(21)US1623002160  US1537001284   US1537001340   US1511000627/ (01)00884838083523(21)US1511000627  US1604001566/ (01)00884838083523(21)US1604001566  US1533001097/ (01)00884838083523(21)US1533001097  US1251000070   US1419000014   US1609001710   US1626002377   US1537001279   US1537001294   US1537001312   US1626002382   US1626002436   US1523000882   US1523000836   US1203010387   US1443000184   US1505000522   US1352000065/ (01)00884838083523(21)US1352000065  US1533001147   US1619002098   US1623002170   US1517000706/ (01)00884838083523(21)US1517000706  US1537001297   US1450000316   US1502000419    US1419000044/ (01)00884838083523(21)US1419000044  US1610001769/ (01)00884838083523(21)US1610001769  US1625002302/ (01)00884838083523(21)US1625002302  US1407000009   US1438000118   US1547001419   US1450000330/ (01)00884838083523(21)US1450000330  US1417000003/ (01)00884838083523(21)US1417000003  US1505000553/ (01)00884838083523(21)US1505000553  US1537001287/ (01)00884838083523(21)US1537001287  US1604001618   US1623002140   US1203010332   US1352000021   US1446000263/ (01)00884838083523(21)US1446000263  US1518000805/ (01)00884838083523(21)US1518000805  US1511000629   US1443000192   US1537001306/ (01)00884838083523(21)US1537001306  US1617002012/ (01)00884838083523(21)US1617002012  US1438000113   US1417000011   US1239000047/ (01)00884838083523(21)US1239000047  US1603001509/ (01)00884838083523(21)US1603001509  US1609001745/ (01)00884838083523(21)US1609001745  US1638002575/ (01)00884838083523(21)US1638002575  US1638002591/ (01)00884838083523(21)US1638002591  US1644002662/ (01)00884838083523(21)US1644002662  US1644002668/ (01)00884838083523(21)US1644002668  US1644002682/ (01)00884838083523(21)US1644002682  US1644002685/ (01)00884838083523(21)US1644002685  US1546001353   US1523000858/ (01)00884838083523(21)US1523000858  US1547001411/ (01)00884838083523(21)US1547001411  US1626002429/ (01)00884838083523(21)US1626002429  US1424000022   US1547001374   US1547001394   US1547001395   US1201010146   US1202010179   US1244000031   US1244000058   US1505000520   US1517000739   US1407000004/ (01)00884838083523(21)US1407000004  US1443000182/ (01)00884838083523(21)US1443000182  US1505000552/ (01)00884838083523(21)US1505000552  US1604001568/ (01)00884838083523(21)US1604001568  US1609001735/ (01)00884838083523(21)US1609001735  US1611001878/ (01)00884838083523(21)US1611001878  US1611001892/ (01)00884838083523(21)US1611001892  US1644002701/ (01)00884838083523(21)US1644002701  US1604001624   US1638002619/ (01)00884838084926(21)US1638002619  US1644002625/ (01)00884838084926(21)US1644002625  US1644002635/ (01)00884838084926(21)US1644002635  US1644002684/ (01)00884838084926(21)US1644002684  US1441000153/ (01)00884838083523(21)US1441000153  US1738002727/ (01)00884838084926(21)US1738002727  US1738002739/ (01)00884838084926(21)US1738002739  US1738002743/ (01)00884838084926(21)US1738002743  US1534001213   US1609001737   US1644002697/ (01)00884838083523(21)US1644002697  US1738002736/ (01)00884838083523(21)US1738002736  US1738002749/ (01)00884838083523(21)US1738002749  US1738002756/ (01)00884838083523(21)US1738002756  US1738002759/ (01)00884838083523(21)US1738002759  US1644002629/ (01)00884838083530(21)US1644002629  US1738002733/ (01)00884838083523(21)US1738002733  US1738002751/ (01)00884838083523(21)US1738002751  US1638002570/ (01)00884838084926(21)US1638002570  US1644002683/ (01)00884838084926(21)US1644002683  US1644002698/ (01)00884838084926(21)US1644002698  US1438000122/ (01)00884838083523(21)US1438000122  US1623002189   US1623002192   US1644002634/ (01)00884838083523(21)US1644002634  US1644002644/ (01)00884838083523(21)US1644002644  US1343000038/ (01)00884838083523(21)US1343000038  US1517000675/ (01)00884838083523(21)US1517000675  US1518000800/ (01)00884838083523(21)US1518000800  US1537001309/ (01)00884838083523(21)US1537001309  US1234000017   US1424000039   US1604001612   US1505000534/ (01)00884838083523(21)US1505000534  US1638002593/ (01)00884838083523(21)US1638002593  US1638002622/ (01)00884838083523(21)US1638002622  US1644002641/ (01)00884838083523(21)US1644002641  US1644002693/ (01)00884838083523(21)US1644002693  US1638002569/ (01)00884838083523(21)US1638002569  US1644002653/ (01)00884838083523(21)US1644002653  US1644002666/ (01)00884838083523(21)US1644002666  US1644002695/ (01)00884838083523(21)US1644002695  US1511000617   US1337000047/ (01)00884838083523(21)US1337000047  US1244000014   US1431000020   US1450000362   US1224000042   US1450000355   US1604001585/ (01)00884838083523(21)US1604001585  US1319000071/ (01)00884838083523(21)US1319000071  US1407000007/ (01)00884838083523(21)US1407000007  US1505000521   US1623002172   US1424000044/ (01)00884838083523(21)US1424000044  US1518000787   US1518000762/ (01)00884838083523(21)US1518000762  US1234000018/ (01)00884838083523(21)US1234000018  US1407000019/ (01)00884838083523(21)US1407000019  US1529001058/ (01)00884838083523(21)US1529001058  US1533001136/ (01)00884838083523(21)US1533001136  US1334000016   US1408000049   US1446000282/ (01)00884838083523(21)US1446000282  US1337000072   US1517000702   US1511000631/ (01)00884838083523(21)US1511000631  US1529001008/ (01)00884838083523(21)US1529001008  US1611001899/ (01)00884838083523(21)US1611001899  US1304000016   US1625002246   US1625002236   US1625002269   US1625002276   US1626002463/ (01)00884838083523(21)US1626002463  US1407000023/ (01)00884838083523(21)US1407000023  US1529000998   US1611001880   US1419000055/ (01)00884838083523(21)US1419000055  US1443000212/ (01)00884838083523(21)US1443000212  US1107010031/ (01)00884838083523(21)US1107010031  US1517000670   US1518000753/ (01)00884838083523(21)US1518000753  US1502000389   US1433000044   US1408000014/ (01)00884838083523(21)US1408000014  US1534001162   US1244000062   US1433000041   US1505000544   US1609001717/ (01)00884838083523(21)US1609001717  US1334000009   US1202010220   US1625002228   US1604001610/ (01)00884838083523(21)US1604001610  US1523000842   US1511000597   US1610001768   US1407000042   US1533001124/ (01)00884838083523(21)US1533001124  US1419000024   US1408000033   US1610001747/ (01)00884838083523(21)US1610001747  US1529001076   US1511000612/ (01)00884838083523(21)US1511000612  US1518000776/ (01)00884838083523(21)US1518000776  US1547001413   US1604001609   US1511000622/ (01)00884838083523(21)US1511000622  US1550001462   US1623002148   US1624002224/ (01)00884838083523(21)US1624002224  US1419000033   US1518000752   US1523000861   US1625002322   US1352000031   US1505000485/ (01)00884838083523(21)US1505000485  US1608001661/ (01)00884838083523(21)US1608001661  US1529001048   US1517000657/ (01)00884838083523(21)US1517000657  US1511000609/ (01)00884838083523(21)US1511000609  US1231000066   US1408000001   US1433000080   US1438000133   US1505000547   US1511000575   US1525000947   US1547001360/ (01)00884838083523(21)US1547001360  US1443000187   US1433000059/ (01)00884838083523(21)US1433000059  US1433000059/ (01)00884838083523(21)US1518000772  US1203010306/ (01)00884838083523(21)US1203010306  US1502000416/ (01)00884838083523(21)US1502000416  US1517000692/ (01)00884838083523(21)US1517000692  US1537001248/ (01)00884838083523(21)US1537001248  US1609001728/ (01)00884838083523(21)US1609001728  US1502000430/ (01)00884838083523(21)US1502000430  US1431000012   US1511000582   US1529001011   US1623002136   US1343000081/ (01)00884838083523(21)US1343000081  US1450000327/ (01)00884838083523(21)US1450000327  US1525000954/ (01)00884838083523(21)US1525000954  US1443000216   US1523000890   US1523000898/ (01)00884838083523(21)US1523000898  US1408000007   US1505000504   US1529001053/ (01)00884838083523(21)US1529001053  US1443000210   US1611001862/ (01)00884838083523(21)US1611001862  US1529000994/ (01)00884838083523(21)US1529000994  US1625002317/ (01)00884838083523(21)US1625002317  US1244000044/ (01)00884838083523(21)US1244000044  US1523000840/ (01)00884838083523(21)US1523000840  US1424000051/ (01)00884838083523(21)US1424000051  US1518000779/ (01)00884838083523(21)US1518000779  US1547001386/ (01)00884838083523(21)US1547001386  US1304000006  US1412000010  US1431000004  US1443000170  US1450000308  US1505000526  US1511000600  US1611001911  US1419000046/ (01)00884838083523(21)US1419000046  US1626002467/ (01)00884838083523(21)US1626002467  US1638002482/ (01)00884838083523(21)US1638002482  US1638002484/ (01)00884838083523(21)US1638002484  US1638002487/ (01)00884838083523(21)US1638002487  US1638002496/ (01)00884838083523(21)US1638002496  US1638002503/ (01)00884838083523(21)US1638002503  US1638002509/ (01)00884838083523(21)US1638002509  US1638002512/ (01)00884838083523(21)US1638002512  US1638002513/ (01)00884838083523(21)US1638002513  US1638002516/ (01)00884838083523(21)US1638002516  US1638002517/ (01)00884838083523(21)US1638002517  US1638002519/ (01)00884838083523(21)US1638002519  US1638002520/ (01)00884838083523(21)US1638002520  US1638002522/ (01)00884838083523(21)US1638002522  US1638002526/ (01 )00884838083523(21)US1638002526  US1638002527/ (01)00884838083523(21)US1638002527  US1638002530/ (01)00884838083523(21)US1638002530  US1638002531/ (01)00884838083523(21)US1638002531  US1638002532/ (01)00884838083523(21)US1638002532  US1638002535/ (01)00884838083523(21)US1638002535  US1638002536/ (01)00884838083523(21)US1638002536  US1638002538/ (01)00884838083523(21)US1638002538  US1638002540/ (01)00884838083523(21)US1638002540  US1638002541/ (01)00884838083523(21)US1638002541  US1638002542/ (01)00884838083523(21)US1638002542  US1638002544/ (01)00884838083523(21)US1638002544  US1638002546/ (01)00884838083523(21)US1638002546  US1638002552/ (01)00884838083523(21)US1638002552  US1638002553/ (01)00884838083523(21)US1638002553  US1638002555/ (01)00884838083523(21)US1638002555  US1638002557/ (01)00884838083523(21)US1638002557  US1638002564/ (01)00884838083523(21)US1638002564  US1638002605/ (01)00884838083523(21)US1638002605  US1638002608/ (01)00884838083523(21)US1638002608  US1638002609/ (01)00884838083523(21)US1638002609  US1638002611/ (01)00884838083523(21)US1638002611  US1638002613/ (01)00884838083523(21)US1638002613  US1638002621/ (01)00884838083523(21)US1638002621  US1644002637/ (01)00884838083523(21)US1644002637  US1644002639/ (01)00884838083523(21)US1644002639  US1644002640/ (01)00884838083523(21)US1644002640  US1644002645/ (01)00884838083523(21)US1644002645  US1644002646/ (01)00884838083523(21)US1644002646  US1644002652/ (01)00884838083523(21)US1644002652  US1644002654/ (01)00884838083523(21)US1644002654  US1644002656/ (01)00884838083523(21)US1644002656  US1644002657/ (01)00884838083523(21)US1644002657  US1644002658/ (01)00884838083523(21)US1644002658  US1644002659/ (01)00884838083523(21)US1644002659  US1644002660/ (01)00884838083523(21)US1644002660  US1644002661/ (01)00884838083523(21)US1644002661  US1644002663/ (01)00884838083523(21)US1644002663  US1644002665/ (01)00884838083523(21)US1644002665  US1644002667/ (01)00884838083523(21)US1644002667  US1644002671/ (01)00884838083523(21)US1644002671  US1644002672/ (01)00884838083523(21)US1644002672  US1644002673/ (01)00884838083523(21)US1644002673  US1644002676/ (01)00884838083523(21)US1644002676  US1644002677/ (01)00884838083523(21)US1644002677  US1644002686/ (01)00884838083523(21)US1644002686  US1644002687/ (01)00884838083523(21)US1644002687  US1644002688/ (01)00884838083523(21)US1644002688  US1644002690/ (01)00884838083523(21)US1644002690  US1644002692/ (01)00884838083523(21)US1644002692  US1644002694/ (01)00884838083523(21)US1644002694  US1644002700/ (01)00884838083523(21)US1644002700  US1343000082  US1352000049  US1407000005  US1407000008  US1407000011  US1407000017  US1407000018  US1407000025  US1407000028  US1407000032  US1407000033  US1407000035  US1407000043  US1407000044  US1407000045  US1407000057  US1407000062  US1408000004  US1408000009  US1408000012  US1408000017  US1408000019  US1408000029  US1408000038  US1408000039  US1408000040  US1408000042  US1408000043  US1408000045  US1408000046  US1408000047  US1412000001  US1412000002  US1412000004  US1412000007  US1412000014  US1412000015  US1412000017  US1412000018  US1412000022  US1412000023  US1412000028  US1412000029  US1412000034  US1412000035  US1412000039  US1412000040  US1412000047  US1412000050  US1417000001  US1417000033  US1417000039  US1419000003  US1419000008  US1419000013  US1419000016  US1419000022  US1419000034  US1419000045  US1419000048  US1419000054  US1419000057  US1419000060  US1424000006  US1424000011  US1424000016  US1424000019  US1424000026  US1424000029  US1424000030  US1424000031  US1424000034  US1424000038  US1424000045  US1424000053  US1424000055  US1431000002  US1431000007  US1431000008  US1431000009  US1431000010  US1431000015  US1431000018  US1431000019  US1431000021  US1431000022  US1431000030  US1431000031  US1431000032  US1431000033  US1431000036  US1431000039  US1433000048  US1433000051  US1433000055  US1433000056  US1433000057  US1433000058  US1433000060  US1433000061  US1433000063  US1433000069  US1433000071  US1433000075  US1433000077  US1433000078  US1433000082  US1433000084  US1433000089  US1433000090  US1438000092  US1438000093  US1438000094  US1438000095  US1438000096  US1438000097  US1438000098  US1438000099  US1438000101  US1438000104  US1438000109  US1438000112  US1438000117  US1438000123  US1438000124  US1438000127  US1438000130  US1438000131  US1438000137  US1438000138  US1438000147  US1443000175  US1443000185  US1443000189  US1443000193  US1443000231  US1443000232  US1443000234  US1446000237  US1446000238  US1446000239  US1446000250  US1446000251  US1446000252  US1446000253  US1446000258  US1446000268  US1446000270  US1446000271  US1446000272  US1446000277  US1446000280  US1446000281  US1446000283  US1446000284  US1446000288  US1446000289  US1446000290  US1446000293  US1446000296  US1446000297  US1446000298  US1446000300  US1446000302  US1446000303  US1446000304  US1446000307  US1450000309  US1450000311  US1450000312  US1450000314  US1450000317  US1450000319  US1450000321  US1450000324  US1450000328  US1450000329  US1450000331  US1450000332  US1450000333  US1450000334  US1450000335  US1450000337  US1450000340  US1450000341  US1450000343  US1450000348  US1450000350  US1450000352  US1450000353  US1450000356  US1450000358  US1450000359  US1450000360  US1450000363  US1450000365  US1450000366  US1450000367  US1450000368  US1450000370  US1450000371  US1450000372  US1450000374  US1450000375  US1450000377  US1502000378  US1502000382  US1502000383  US1502000385  US1502000386  US1502000393  US1502000394  US1502000398  US1502000403  US1502000404  US1502000406  US1502000407  US1502000408  US1502000413  US1502000415  US1502000418  US1502000421  US1502000425  US1502000438  US1502000456  US1502000457  US1502000458  US1502000468  US1502000469  US1502000476  US1505000495  US1505000539  US1511000594  US1511000601  US1511000606  US1511000607  US1511000625  US1511000628  US1511000632  US1511000636  US1511000641  US1517000653  US1517000664  US1517000718  US1517000720  US1517000721  US1517000722  US1517000723  US1517000728  US1518000748  US1518000749  US1518000750  US1518000754  US1518000757  US1518000759  US1518000760  US1518000767  US1518000769  US1518000799  US1523000808  US1523000814  US1523000818  US1523000819  US1523000821  US1523000824  US1523000826  US1523000829  US1523000830  US1523000831  US1523000832  US1523000834  US1523000835  US1523000838  US1523000843  US1523000845  US1523000846  US1523000847  US1523000850  US1523000851  US1523000852  US1523000853  US1523000856  US1523000857  US1523000860  US1523000865  US1523000868  US1523000871  US1523000874  US1523000876  US1523000877  US1523000878  US1523000880  US1523000899  US1525000913  US1525000918  US1525000928  US1525000932  US1525000937  US1525000948  US1525000951  US1525000953  US1525000957  US1525000965  US1525000967  US1525000969  US1525000971  US1525000977  US1525000980  US1525000981  US1525000982  US1525000983  US1529000996  US1529001002  US1529001003  US1529001007  US1529001009  US1529001015  US1529001018  US1529001020  US1529001022  US1529001023  US1529001026  US1529001030  US1529001034  US1529001035  US1529001039  US1529001040  US1529001041  US1529001043  US1529001060  US1529001061  US1529001062  US1529001065  US1529001066  US1529001069  US1529001070  US1529001071  US1529001074  US1529001091  US1533001101  US1533001105  US1533001107  US1533001108  US1533001109  US1533001112  US1533001113  US1533001120  US1533001122  US1533001123  US1533001131  US1533001135  US1533001138  US1533001141  US1533001142  US1533001143  US1533001145  US1533001149  US1534001152  US1534001154  US1534001156  US1534001161  US1534001176  US1534001178  US1534001191  US1534001197  US1534001198  US1534001200  US1534001206  US1534001207  US1534001219  US1534001222  US1534001225  US1534001234  US1534001239  US1534001244  US1537001249  US1537001250  US1537001251  US1537001252  US1537001253  US1537001254  US1537001255  US1537001256  US1537001257  US1537001262  US1537001264  US1537001267  US1537001269  US1537001270  US1537001271  US1537001274  US1537001278  US1537001285  US1537001290  US1537001291  US1537001292  US1537001304  US1537001313  US1537001319  US1537001322  US1537001326  US1537001327  US1537001328  US1537001329  US1537001332  US1537001333  US1537001334  US1537001335  US1537001336  US1537001337  US1537001339  US1546001352  US1546001354  US1547001359  US1547001361  US1547001363  US1547001375  US1547001382  US1547001418  US1547001422  US1547001424  US1547001425  US1547001428  US1547001434  US1547001436  US1550001444  US1550001452  US1550001456  US1550001458  US1550001459  US1550001461  US1550001481  US1550001490  US1550001492  US1550001496  US1550001497  US1550001499  US1603001507  US1603001513  US1603001517  US1603001523  US1603001524  US1603001534  US1603001535  US1603001537  US1603001540  US1603001542  US1603001543  US1603001544  US1603001545  US1603001546  US1603001548  US1603001553  US1603001554  US1603001558  US1603001559  US1603001560  US1603001561  US1603001562  US1603001563  US1603001564  US1604001567  US1604001577  US1604001579  US1604001587  US1604001589  US1608001629  US1608001660  US1608001667  US1608001674  US1608001681  US1609001686  US1609001694  US1609001711  US1609001713  US1609001720  US1609001721  US1609001730  US1609001736  US1609001740  US1609001741  US1609001742  US1609001744  US1610001746  US1610001751  US1610001752  US1610001754  US1610001756  US1610001758  US1610001759  US1610001763  US1610001765  US1610001766  US1610001767  US1610001770  US1610001777  US1610001784  US1610001790  US1610001795  US1610001799  US1610001800  US1610001802  US1611001810  US1611001812  US1611001813  US1611001818  US1611001822  US1611001823  US1611001824  US1611001825  US1611001827  US1611001830  US1611001833  US1611001836  US1611001837  US1611001840  US1611001842  US1611001849  US1611001851  US1611001853  US1611001857  US1611001859  US1611001861  US1611001864  US1611001865  US1611001866  US1611001873  US1611001874  US1611001879  US1611001881  US1611001884  US1611001887  US1611001894  US1611001898  US1611001903  US1611001905  US1611001907  US1611001912  US1611001915  US1611001921  US1613001939  US1613001941  US1613001942  US1613001946  US1617002031  US1619002046  US1619002052  US1619002053  US1619002056  US1619002063  US1619002065  US1619002066  US1619002069  US1619002072  US1619002073  US1619002074  US1619002075  US1619002086  US1619002087  US1619002088  US1619002089  US1619002091  US1619002092  US1619002093  US1619002094  US1619002095  US1619002096  US1619002100  US1619002101  US1623002106  US1623002111  US1623002130  US1623002135  US1623002138  US1623002142  US1623002145  US1623002164  US1623002165  US1623002169  US1623002174  US1623002177  US1623002181  US1623002185  US1623002187  US1623002188  US1623002190  US1623002195  US1624002200  US1624002204  US1624002215  US1624002220  US1625002229  US1625002235  US1625002248  US1625002252  US1625002254  US1625002258  US1625002265  US1625002270  US1625002283  US1625002285  US1625002288  US1625002296  US1625002299  US1625002300  US1625002301  US1625002303  US1625002304  US1625002305  US1625002306  US1625002308  US1625002309  US1625002310  US1625002313  US1625002314  US1625002316  US1625002319  US1625002321  US1625002323  US1625002325  US1625002326  US1625002327  US1625002328  US1625002329  US1625002331  US1625002332  US1625002334  US1625002335  US1625002336  US1625002337  US1625002339  US1625002341  US1625002342  US1625002343  US1626002395  US1626002400  US1626002414  US1626002425  US1626002428  US1626002437  US1626002440  US1626002441  US1626002442  US1626002445  US1626002446  US1626002448  US1626002449  US1626002450  US1626002452  US1626002453  US1626002456  US1638002539/ (01)00884838083523(21)US1638002539  US1407000001   US1407000061   US1412000019   US1412000024   US1412000031   US1424000001   US1424000028   US1431000006   US1433000073   US1438000102   US1438000114   US1443000179   US1446000244   US1446000275   US1446000306   US1450000315   US1450000339   US1502000400   US1502000410   US1502000461   US1511000630   US1523000879   US1523000891   US1525000902   US1525000931   US1537001282   US1537001317   US1550001493   US1608001668   US1608001676   US1611001811   US1611001843   US1611001854   US1611001890   US1619002051   US1625002280   US1625002295   US1450000310   US1352000004   US1523000841   US1248000001/ (01)00884838083523(21)US1248000001  US1109010093   US1109010108   US1534001155   US1534001172   US1534001236   US1611001841   US1407000010   US1417000012   US1533001139/ (01)00884838083523(21)US1533001139  US1505000483   US1239000040   US1412000006   US1424000015   US1511000595   US1352000014   US1352000008   US1537001320   US1505000505   US1319000087/ (01)00884838083523(21)US1319000087  US1407000015/ (01)00884838083523(21)US1407000015  US1603001552   US1518000761/ (01)00884838083523(21)US1518000761  US1537001289/ (01)00884838083523(21)US1537001289  US1337000053   US1505000512   US1523000833/ (01)00884838083523(21)US1523000833  US1625002307   US1511000583   US1537001316/ (01)00884838083523(21)US1537001316  US1523000875/ (01)00884838083523(21)US1523000875  US1534001228/ (01)00884838083523(21)US1534001228  US1611001889/ (01)00884838083523(21)US1611001889  US1523000822/ (01)00884838083523(21)US1523000822  US1511000621  US1534001169/ (01)00884838083523(21)US1534001169  US1352000019   US1234000077   US1446000246   US1529001017   US1502000405/ (01)00884838083523(21)US1502000405  US1610001776   US1547001431/ (01)00884838083523(21)US1547001431  US1352000064   US1511000576   US1626002392/ (01)00884838083523(21)US1626002392  US1625002289   US1244000024/ (01)00884838083523(21)US1244000024  US1511000619   US1624002201   US1523000895/ (01)00884838083523(21)US1523000895  US1529000995   US1638002620/ (01)00884838082113(21)US1638002620  US1419000051   US1511000616/ (01)00884838083523(21)US1511000616  US1604001591   US1623002139   US1446000265   US1424000020   US1523000892   US1446000292/ (01)00884838083523(21)US1446000292  US1529001025   US1603001556   US1511000610/ (01)00884838083523(21)US1511000610  US1251000074   US1441000155   US1517000688/ (01)00884838083523(21)US1517000688  US1604001611/ (01)00884838083523(21)US1604001611  US1609001689/ (01)00884838083523(21)US1609001689  US1610001761/ (01)00884838083523(21)US1610001761  US1419000010   US1419000007   US1319000058   US1611001904/ (01)00884838083523(21)US1611001904  US1419000049   US1338000004   US1547001369/ (01)00884838083523(21)US1547001369  US1523000844/ (01)00884838083523(21)US1523000844  US1424000003   US1529001038/ (01)00884838083523(21)US1529001038  US1623002162   US1523000823/ (01)00884838083523(21)US1523000823  US1626002462/ (01)00884838083523(21)US1626002462  US1446000236   US1502000411   US1517000736/ (01)00884838083523(21)US1517000736  US1604001607/ (01)00884838083523(21)US1604001607  US1343000067   US1523000862       Model Number:  453564621791  Serial Numbers/UDI (if available):  US1734000898, (01)00884838084933(21)US1734000898  US1734000907, (01)00884838084933(21)US1734000907  US1734000909, (01)00884838084933(21)US1734000909  US1729000608, (01)00884838084902(21)US1729000608  US1731000696, (01)00884838084902(21)US1731000696  US1736000924, (01)00884838084902(21)US1736000924  US1736000933, (01)00884838084902(21)US1736000933  US1736000936, (01)00884838084902(21)US1736000936  US1736000939, (01)00884838084902(21)US1736000939  US1736000942, (01)00884838084902(21)US1736000942  US1736000956, (01)00884838084902(21)US1736000956  US1736000959, (01)00884838084902(21)US1736000959  US1736000962, (01)00884838084902(21)US1736000962  US1736000965, (01)00884838084902(21)US1736000965  US1736000967, (01)00884838084902(21)US1736000967  US1736000971, (01)00884838084902(21)US1736000971  US1736000973, (01)00884838084902(21)US1736000973  US1736000977, (01)00884838084902(21)US1736000977  US1736000998, (01)00884838084902(21)US1736000998  US1736001000, (01)00884838084902(21)US1736001000  US1736001004, (01)00884838084902(21)US1736001004  US1736001006, (01)00884838084902(21)US1736001006  US1736001009, (01)00884838084902(21)US1736001009  US1738001012, (01)00884838084902(21)US1738001012  US1738001058, (01)00884838084902(21)US1738001058  US1738001083, (01)00884838084902(21)US1738001083  US1738001085, (01)00884838084902(21)US1738001085  US1738001095, (01)00884838084902(21)US1738001095  US1746001211, (01)00884838084902(21)US1746001211  US1746001213, (01)00884838084902(21)US1746001213  US1746001237, (01)00884838084902(21)US1746001237  US1746001238, (01)00884838084902(21)US1746001238  US1724000514, (01)00884838084902(21)US1724000514  US1729000605, (01)00884838084902(21)US1729000605  US1729000631, (01)00884838084902(21)US1729000631  US1729000652, (01)00884838084902(21)US1729000652  US1731000702, (01)00884838084902(21)US1731000702  US1731000740, (01)00884838084902(21)US1731000740  US1734000790, (01)00884838084902(21)US1734000790  US1734000801, (01)00884838084902(21)US1734000801  US1734000863, (01)00884838084902(21)US1734000863  US1734000882, (01)00884838084902(21)US1734000882  US1734000888, (01)00884838084902(21)US1734000888  US1736000944, (01)00884838084902(21)US1736000944  US1736000983, (01)00884838084902(21)US1736000983  US1736000985, (01)00884838084902(21)US1736000985  US1736000987, (01)00884838084902(21)US1736000987  US1736000991, (01)00884838084902(21)US1736000991  US1736001003, (01)00884838084902(21)US1736001003  US1738001029, (01)00884838084902(21)US1738001029  US1738001120, (01)00884838084902(21)US1738001120  US1734000850, (01)00884838083523(21)US1734000850  US1734000899, (01)00884838083523(21)US1734000899  US1734000908, (01)00884838083523(21)US1734000908  US1736000919, (01)00884838083523(21)US1736000919  US1736000925, (01)00884838083523(21)US1736000925  US1738001020, (01)00884838083523(21)US1738001020  US1709000217, (01)00884838083523(21)US1709000217  US1716000288, (01)00884838083523(21)US1716000288  US1716000293, (01)00884838083523(21)US1716000293  US1716000306, (01)00884838083523(21)US1716000306  US1716000308, (01)00884838083523(21)US1716000308  US1716000312, (01)00884838083523(21)US1716000312  US1716000322, (01)00884838083523(21)US1716000322  US1716000342, (01)00884838083523(21)US1716000342  US1738001060, (01)00884838084933(21)US1738001060  US1738001079, (01)00884838084933(21)US1738001079  US1731000745, (01)00884838084933(21)US1731000745  US1731000753, (01)00884838084933(21)US1731000753  US1734000864, (01)00884838084933(21)US1734000864  US1738001118, (01)00884838084933(21)US1738001118  US1738001124, (01)00884838084933(21)US1738001124  US1729000646, (01)00884838084902(21)US1729000646  US1729000623, (01)00884838084926(21)US1729000623  US1729000640, (01)00884838084926(21)US1729000640  US1724000503, (01)00884838084926(21)US1724000503  US1724000511, (01)00884838084926(21)US1724000511  US1724000545, (01)00884838084926(21)US1724000545  US1724000547, (01)00884838084926(21)US1724000547  US1729000571, (01)00884838083523(21)US1729000571  US1729000616, (01)00884838083523(21)US1729000616  US1729000619, (01)00884838083523(21)US1729000619  US1734000816, (01)00884838083523(21)US1734000816  US1734000817, (01)00884838083523(21)US1734000817  US1729000632, (01)00884838084926(21)US1729000632  US1729000575, (01)00884838084926(21)US1729000575  US1731000677, (01)00884838084926(21)US1731000677  US1731000664, (01)00884838084902(21)US1731000664  US1731000732, (01)00884838083523(21)US1731000732  US1729000612, (01)00884838084926(21)US1729000612  US1731000685, (01)00884838084902(21)US1731000685  US1652000033, (01)00884838083523(21)US1652000033  US1724000470, (01)00884838083523(21)US1724000470  US1729000576, (01)00884838083523(21)US1729000576  US1729000565, (01)00884838084902(21)US1729000565  US1729000618, (01)00884838083523(21)US1729000618  US1731000682, (01)00884838083523(21)US1731000682  US1724000490, (01)00884838084902(21)US1724000490  US1734000792, (01)00884838083523(21)US1734000792  US1734000825, (01)00884838083523(21)US1734000825  US1738001133, (01)00884838084926(21)US1738001133  US1746001143, (01)00884838084926(21)US1746001143  US1746001181, (01)00884838084926(21)US1746001181  US1720000350, (01)00884838083523(21)US1720000350  US1720000365, (01)00884838083523(21)US1720000365  US1729000638, (01)00884838084902(21)US1729000638  US1652000021, (01)00884838083523(21)US1652000021  US1652000057, (01)00884838083523(21)US1652000057  US1652000071, (01)00884838083523(21)US1652000071  US1734000877, (01)00884838084902(21)US1734000877  US1738001077, (01)00884838084902(21)US1738001077  US1729000649, (01)00884838083523(21)US1729000649  US1731000675, (01)00884838083523(21)US1731000675  US1720000459, (01)00884838084926(21)US1720000459  US1709000206, (01)00884838083523(21)US1709000206  US1709000223, (01)00884838083523(21)US1709000223  US1734000866, (01)00884838083523(21)US1734000866  US1744001138, (01)00884838083523(21)US1744001138  US1709000207, (01)00884838083523(21)US1709000207  US1709000254, (01)00884838083523(21)US1709000254  US1729000609, (01)00884838084902(21)US1729000609  US1716000297, (01)00884838084926(21)US1716000297  US1716000326, (01)00884838084926(21)US1716000326  US1720000373, (01)00884838084926(21)US1720000373  US1734000879, (01)00884838083523(21)US1734000879  US1738001125, (01)00884838083523(21)US1738001125  US1738001129, (01)00884838083523(21)US1738001129  US1709000214, (01)00884838083523(21)US1709000214  US1709000232, (01)00884838083523(21)US1709000232  US1734000791, (01)00884838084933(21)US1734000791  US1734000830, (01)00884838084933(21)US1734000830  US1736000923, (01)00884838084933(21)US1736000923  US1729000630, (01)00884838084902(21)US1729000630  US1738001104, (01)00884838084926(21)US1738001104  US1746001193, (01)00884838084926(21)US1746001193  US1746001196, (01)00884838084926(21)US1746001196  US1746001197, (01)00884838084926(21)US1746001197  US1724000505, (01)00884838084902(21)US1724000505  US1734000871, (01)00884838084902(21)US1734000871  US1720000387, (01)00884838084926(21)US1720000387  US1720000389, (01)00884838084926(21)US1720000389  US1720000425, (01)00884838084926(21)US1720000425  US1724000541, (01)00884838083530(21)US1724000541  US1708000193, (01)00884838083523(21)US1708000193  US1716000289, (01)00884838083523(21)US1716000289  US1716000299, (01)00884838083523(21)US1716000299  US1709000210, (01)00884838083523(21)US1709000210  US1709000226, (01)00884838083523(21)US1709000226  US1734000775, (01)00884838083523(21)US1734000775  US1734000776, (01)00884838083523(21)US1734000776  US1734000780, (01)00884838083523(21)US1734000780  US1734000783, (01)00884838083523(21)US1734000783  US1734000785, (01)00884838083523(21)US1734000785  US1734000786, (01)00884838083523(21)US1734000786  US1734000807, (01)00884838083523(21)US1734000807  US1734000814, (01)00884838083523(21)US1734000814  US1734000836, (01)00884838083523(21)US1734000836  US1734000852, (01)00884838083523(21)US1734000852  US1734000870, (01)00884838083523(21)US1734000870  US1734000878, (01)00884838083523(21)US1734000878  US1724000554, (01)00884838083523(21)US1724000554  US1729000585, (01)00884838083523(21)US1729000585  US1729000614, (01)00884838083523(21)US1729000614  US1729000639, (01)00884838083523(21)US1729000639  US1709000287, (01)00884838083523(21)US1709000287  US1716000294, (01)00884838083523(21)US1716000294  US1716000332, (01)00884838083523(21)US1716000332  US1716000335, (01)00884838083523(21)US1716000335  US1716000339, (01)00884838083523(21)US1716000339  US1724000518, (01)00884838084926(21)US1724000518  US1724000524, (01)00884838084926(21)US1724000524  US1652000006, (01)00884838083530(21)US1652000006  US1652000058, (01)00884838083530(21)US1652000058  US1652000073, (01)00884838083530(21)US1652000073  US1652000069, (01)00884838083530(21)US1652000069  US1652000076, (01)00884838083530(21)US1652000076  US1652000094, (01)00884838083530(21)US1652000094  US1709000213, (01)00884838084926(21)US1709000213  US1736000979, (01)00884838084902(21)US1736000979  US1709000234, (01)00884838084926(21)US1709000234  US1720000360, (01)00884838084926(21)US1720000360  US1720000367, (01)00884838084926(21)US1720000367  US1720000372, (01)00884838084926(21)US1720000372  US1716000311, (01)00884838084926(21)US1716000311  US1716000323, (01)00884838084926(21)US1716000323  US1734000857, (01)00884838083523(21)US1734000857  US1734000862, (01)00884838083523(21)US1734000862  US1709000259, (01)00884838083523(21)US1709000259  US1652000020, (01)00884838084926(21)US1652000020  US1652000038, (01)00884838084926(21)US1652000038  US1734000894, (01)00884838083523(21)US1734000894  US1734000900, (01)00884838083523(21)US1734000900  US1734000901, (01)00884838083523(21)US1734000901  US1738001031, (01)00884838083523(21)US1738001031  US1738001046, (01)00884838083523(21)US1738001046  US1738001059, (01)00884838083523(21)US1738001059  US1738001072, (01)00884838083523(21)US1738001072  US1738001086, (01)00884838083523(21)US1738001086  US1731000703, (01)00884838084902(21)US1731000703  US1729000586, (01)00884838084902(21)US1729000586  US1724000478, (01)00884838084933(21)US1724000478  US1724000484, (01)00884838084933(21)US1724000484  US1720000374, (01)00884838084926(21)US1720000374  US1720000376, (01)00884838084926(21)US1720000376  US1729000589, (01)00884838083530(21)US1729000589  US1709000220, (01)00884838083523(21)US1709000220  US1709000230, (01)00884838083523(21)US1709000230  US1709000235, (01)00884838083523(21)US1709000235  US1709000237, (01)00884838083523(21)US1709000237  US1731000756, (01)00884838084926(21)US1731000756  US1734000884, (01)00884838084926(21)US1734000884  US1652000010, (01)00884838083523(21)US1652000010  US1708000107, (01)00884838083523(21)US1708000107  US1708000133, (01)00884838083523(21)US1708000133  US1708000142, (01)00884838083523(21)US1708000142  US1709000248, (01)00884838084902(21)US1709000248  US1652000019, (01)00884838083523(21)US1652000019  US1724000515, (01)00884838084933(21)US1724000515  US1724000519, (01)00884838084933(21)US1724000519  US1724000544, (01)00884838084933(21)US1724000544  US1724000553, (01)00884838084933(21)US1724000553  US1720000410, (01)00884838084926(21)US1720000410  US1720000442, (01)00884838084926(21)US1720000442  US1720000447, (01)00884838084926(21)US1720000447  US1720000453, (01)00884838084926(21)US1720000453  US1724000525, (01)00884838084933(21)US1724000525  US1724000533, (01)00884838084933(21)US1724000533  US1720000384, (01)00884838084933(21)US1720000384  US1720000391, (01)00884838084933(21)US1720000391  US1720000400, (01)00884838084933(21)US1720000400  US1720000412, (01)00884838084933(21)US1720000412  US1734000847, (01)00884838084933(21)US1734000847  US1734000849, (01)00884838084933(21)US1734000849  US1734000911, (01)00884838084933(21)US1734000911  US1736000915, (01)00884838084933(21)US1736000915  US1736000917, (01)00884838084933(21)US1736000917  US1708000119, (01)00884838083523(21)US1708000119  US1708000135, (01)00884838083523(21)US1708000135  US1708000138, (01)00884838083523(21)US1708000138  US1708000145, (01)00884838083523(21)US1708000145  US1708000159, (01)00884838083523(21)US1708000159  US1708000161, (01)00884838083523(21)US1708000161  US1729000655, (01)00884838083523(21)US1729000655  US1731000666, (01)00884838083523(21)US1731000666  US1731000727, (01)00884838083523(21)US1731000727  US1731000737, (01)00884838083523(21)US1731000737  US1731000746, (01)00884838083523(21)US1731000746  US1731000747, (01)00884838083523(21)US1731000747  US1731000750, (01)00884838083523(21)US1731000750  US1731000752, (01)00884838083523(21)US1731000752  US1734000768, (01)00884838083523(21)US1734000768  US1731000665, (01)00884838084902(21)US1731000665  US1652000027, (01)00884838083523(21)US1652000027  US1652000041, (01)00884838083523(21)US1652000041  US1652000061, (01)00884838083523(21)US1652000061  US1652000075, (01)00884838083523(21)US1652000075  US1652000081, (01)00884838083523(21)US1652000081  US1652000082, (01)00884838083523(21)US1652000082  US1652000085, (01)00884838083523(21)US1652000085  US1652000088, (01)00884838083523(21)US1652000088  US1652000090, (01)00884838083523(21)US1652000090  US1709000255, (01)00884838083523(21)US1709000255  US1652000014, (01)00884838083523(21)US1652000014  US1652000060, (01)00884838083523(21)US1652000060  US1716000290, (01)00884838084926(21)US1716000290  US1716000324, (01)00884838084926(21)US1716000324  US1729000577, (01)00884838083523(21)US1729000577  US1734000891, (01)00884838084902(21)US1734000891  US1729000567, (01)00884838083523(21)US1729000567  US1729000568, (01)00884838083523(21)US1729000568  US1729000569, (01)00884838083523(21)US1729000569  US1729000578, (01)00884838083523(21)US1729000578  US1729000599, (01)00884838083523(21)US1729000599  US1724000481, (01)00884838084902(21)US1724000481  US1720000455, (01)00884838084926(21)US1720000455  US1724000462, (01)00884838084926(21)US1724000462  US1731000711, (01)00884838084926(21)US1731000711  US1731000722, (01)00884838084926(21)US1731000722  US1652000029, (01)00884838083523(21)US1652000029  US1652000080, (01)00884838083523(21)US1652000080  US1738001055, (01)00884838083523(21)US1738001055  US1738001091, (01)00884838083523(21)US1738001091  US1724000528, (01)00884838084926(21)US1724000528  US1724000529, (01)00884838084926(21)US1724000529  US1734000859, (01)00884838084933(21)US1734000859  US1708000175, (01)00884838084902(21)US1708000175  US1709000264, (01)00884838084902(21)US1709000264  US1724000497, (01)00884838084933(21)US1724000497  US1724000475, (01)00884838083523(21)US1724000475  US1729000590, (01)00884838083523(21)US1729000590  US1729000591, (01)00884838083523(21)US1729000591  US1652000032, (01)00884838083530(21)US1652000032  US1652000051, (01)00884838083530(21)US1652000051  US1734000777, (01)00884838084933(21)US1734000777  US1734000781, (01)00884838084933(21)US1734000781  US1734000808, (01)00884838084933(21)US1734000808  US1729000657, (01)00884838084902(21)US1729000657  US1734000881, (01)00884838084933(21)US1734000881  US1736000961, (01)00884838084933(21)US1736000961  US1652000015, (01)00884838083523(21)US1652000015  US1652000040, (01)00884838083523(21)US1652000040  US1720000348, (01)00884838083523(21)US1720000348  US1720000349, (01)00884838083523(21)US1720000349  US1720000352, (01)00884838083523(21)US1720000352  US1720000417, (01)00884838083523(21)US1720000417  US1724000531, (01)00884838084902(21)US1724000531  US1652000044, (01)00884838083523(21)US1652000044  US1652000070, (01)00884838083523(21)US1652000070  US1652000072, (01)00884838083523(21)US1652000072  US1652000078, (01)00884838083523(21)US1652000078  US1652000079, (01)00884838083523(21)US1652000079  US1652000084, (01)00884838083523(21)US1652000084  US1652000086, (01)00884838083523(21)US1652000086  US1652000087, (01)00884838083523(21)US1652000087  US1652000089, (01)00884838083523(21)US1652000089  US1652000091, (01)00884838083523(21)US1652000091  US1652000092, (01)00884838083523(21)US1652000092  US1652000095, (01)00884838083523(21)US1652000095  US1652000096, (01)00884838083523(21)US1652000096  US1724000502, (01)00884838084926(21)US1724000502  US1724000508, (01)00884838084926(21)US1724000508  US1729000574, (01)00884838083523(21)US1729000574  US1729000584, (01)00884838083523(21)US1729000584  US1729000595, (01)00884838083523(21)US1729000595  US1709000225, (01)00884838084926(21)US1709000225  US1716000316, (01)00884838084926(21)US1716000316  US1720000369, (01)00884838084926(21)US1720000369  US1720000371, (01)00884838084926(21)US1720000371  US1731000734, (01)00884838084926(21)US1731000734  US1734000769, (01)00884838084926(21)US1734000769  US1734000778, (01)00884838083523(21)US1734000778  US1734000789, (01)00884838083523(21)US1734000789  US1734000833, (01)00884838083523(21)US1734000833  US1734000834, (01)00884838083523(21)US1734000834  US1734000838, (01)00884838083523(21)US1734000838  US1734000840, (01)00884838083523(21)US1734000840  US1734000843, (01)00884838083523(21)US1734000843  US1734000846, (01)00884838083523(21)US1734000846  US1734000854, (01)00884838083523(21)US1734000854  US1734000874, (01)00884838083523(21)US1734000874  US1736000964, (01)00884838084902(21)US1736000964  US1709000216, (01)00884838084926(21)US1709000216  US1709000250, (01)00884838084926(21)US1709000250  US1731000691, (01)00884838083523(21)US1731000691  US1731000720, (01)00884838083523(21)US1731000720  US1652000013, (01)00884838083530(21)US1652000013  US1652000077, (01)00884838083530(21)US1652000077  US1652000093, (01)00884838083530(21)US1652000093  US1729000634, (01)00884838084902(21)US1729000634  US1724000500, (01)00884838084902(21)US1724000500  US1729000570, (01)00884838084902(21)US1729000570  US1724000549, (01)00884838084926(21)US1724000549  US1736000966, (01)00884838084926(21)US1736000966  US17380010 94, (01)00884838084926(21)US1738001094  US1738001106, (01)00884838084926(21)US1738001106  US1724000516, (01)00884838084926(21)US1724000516  US1724000534, (01)00884838084926(21)US1724000534  US1724000542, (01)00884838084926(21)US1724000542  US1709000249, (01)00884838083530(21)US1709000249  US1709000275, (01)00884838083530(21)US1709000275  US1720000357, (01)00884838083530(21)US1720000357  US1709000261, (01)00884838083523(21)US1709000261  US1709000284, (01)00884838083523(21)US1709000284  US1729000641, (01)00884838084902(21)US1729000641  US1734000763, (01)00884838084926(21)US1734000763  US1716000295, (01)00884838083523(21)US1716000295  US1716000321, (01)00884838083523(21)US1716000321  US1738001100, (01)00884838084926(21)US1738001100  US1746001142, (01)00884838084926(21)US1746001142  US1746001174, (01)00884838084926(21)US1746001174  US1746001180, (01)00884838084926(21)US1746001180  US1746001186, (01)00884838084926(21)US1746001186  US1729000629, (01)00884838084902(21)US1729000629  US1734000880, (01)00884838084902(21)US1734000880  US1720000356, (01)00884838084926(21)US1720000356  US1720000433, (01)00884838084926(21)US1720000433  US1652000054, (01)00884838083523(21)US1652000054  US1652000059, (01)00884838083523(21)US1652000059  US1652000067, (01)00884838083523(21)US1652000067  US1729000592, (01)00884838084902(21)US1729000592  US1729000596, (01)00884838084902(21)US1729000596  US1738001087, (01)00884838083523(21)US1738001087  US1652000012, (01)00884838083523(21)US1652000012  US1652000017, (01)00884838083523(21)US1652000017  US1652000023, (01)00884838083523(21)US1652000023  US1652000030, (01)00884838083523(21)US1652000030  US1652000048, (01)00884838083523(21)US1652000048  US1652000049, (01)00884838083523(21)US1652000049  US1652000055, (01)00884838083523(21)US1652000055  US1652000056, (01)00884838083523(21)US1652000056  US1652000065, (01)00884838083523(21)US1652000065  US1652000068, (01)00884838083523(21)US1652000068  US1708000103, (01)00884838083523(21)US1708000103  US1708000105, (01)00884838083523(21)US1708000105  US1708000108, (01)00884838083523(21)US1708000108  US1708000109, (01)00884838083523(21)US1708000109  US1708000111, (01)00884838083523(21)US1708000111  US1708000112, (01)00884838083523(21)US1708000112  US1708000117, (01)00884838083523(21)US1708000117  US1708000118, (01)00884838083523(21)US1708000118  US1708000122, (01)00884838083523(21)US1708000122  US1736000992, (01)00884838084902(21)US1736000992  US1708000144, (01)00884838083530(21)US1708000144  US1709000251, (01)00884838083530(21)US1709000251  US1724000540, (01)00884838083530(21)US1724000540  US1709000269, (01)00884838084926(21)US1709000269  US1708000150, (01)00884838083530(21)US1708000150  US1720000361, (01)00884838083530(21)US1720000361  US1720000424, (01)00884838083530(21)US1720000424  US1720000429, (01)00884838083530(21)US1720000429  US1720000432, (01)00884838083530(21)US1720000432  US1652000007, (01)00884838083523(21)US1652000007  US1652000024, (01)00884838083523(21)US1652000024  US1652000036, (01)00884838083523(21)US1652000036  US1652000053, (01)00884838083523(21)US1652000053  US1652000083, (01)00884838083523(21)US1652000083  US1724000522, (01)00884838083523(21)US1724000522  US1729000580, (01)00884838083523(21)US1729000580  US1729000622, (01)00884838083523(21)US1729000622  US1729000651, (01)00884838083523(21)US1729000651  US1709000212, (01)00884838084926(21)US1709000212  US1716000303, (01)00884838084926(21)US1716000303  US1720000366, (01)00884838084926(21)US1720000366  US1720000368, (01)00884838084926(21)US1720000368  US1736000954, (01)00884838083523(21)US1736000954  US1738001057, (01)00884838083523(21)US1738001057  US1738001062, (01)00884838083523(21)US1738001062  US1738001068, (01)00884838083523(21)US1738001068  US1738001090, (01)00884838083523(21)US1738001090  US1738001126, (01)00884838083523(21)US1738001126  US1708000114, (01)00884838083523(21)US1708000114  US1708000127, (01)00884838083523(21)US1708000127  US1708000132, (01)00884838083523(21)US1708000132  US1708000140, (01)00884838083523(21)US1708000140  US1708000146, (01)00884838083523(21)US1708000146  US1708000149, (01)00884838083523(21)US1708000149  US1708000152, (01)00884838083523(21)US1708000152  US1708000154, (01)00884838083523(21)US1708000154  US1708000163, (01)00884838083523(21)US1708000163  US1708000168, (01)00884838083523(21)US1708000168  US1708000170, (01)00884838083523(21)US1708000170  US1708000171, (01)00884838083523(21)US1708000171  US1708000173, (01)00884838083523(21)US1708000173  US1708000174, (01)00884838083523(21)US1708000174  US1708000183, (01)00884838083523(21)US1708000183  US1708000198, (01)00884838083523(21)US1708000198  US1708000200, (01)00884838083523(21)US1708000200  US1708000202, (01)00884838083523(21)US1708000202  US1709000205, (01)00884838083523(21)US1709000205  US1736001010, (01)00884838083523(21)US1736001010  US1738001032, (01)00884838083523(21)US1738001032  US1738001045, (01)00884838083523(21)US1738001045  US1738001049, (01)00884838083523(21)US1738001049  US1738001063, (01)00884838083523(21)US1738001063  US1738001067, (01)00884838083523(21)US1738001067  US1738001074, (01)00884838083523(21)US1738001074  US1738001076, (01)00884838083523(21)US1738001076  US1709000222, (01)00884838083523(21)US1709000222  US1720000370, (01)00884838083523(21)US1720000370  US1729000625, (01)00884838083523(21)US1729000625  US1729000628, (01)00884838083523(21)US1729000628  US1729000636, (01)00884838083523(21)US1729000636  US1729000637, (01)00884838083523(21)US1729000637  US1729000643, (01)00884838083523(21)US1729000643  US1729000644, (01)00884838083523(21)US1729000644  US1729000645, (01)00884838083523(21)US1729000645  US1724000493, (01)00884838084902(21)US1724000493  US1731000662, (01)00884838084902(21)US1731000662  US1709000219, (01)00884838083530(21)US1709000219  US1709000253, (01)00884838083530(21)US1709000253  US1746001185, (01)00884838084926(21)US1746001185  US1709000233, (01)00884838084926(21)US1709000233  US1709000236, (01)00884838084926(21)US1709000236  US1709000239, (01)00884838084926(21)US1709000239  US1709000240, (01)00884838084926(21)US1709000240  US1709000276, (01)00884838084926(21)US1709000276  US1734000867, (01)00884838083523(21)US1734000867  US1731000695, (01)00884838083523(21)US1731000695  US1731000697, (01)00884838083523(21)US1731000697  US1731000708, (01)00884838083523(21)US1731000708  US1731000710, (01)00884838083523(21)US1731000710  US1731000724, (01)00884838083523(21)US1731000724  US1652000031, (01)00884838083530(21)US1652000031  US1652000039, (01)00884838083530(21)US1652000039  US1652000074, (01)00884838083530(21)US1652000074  US1738001066, (01)00884838083523(21)US1738001066  US1734000893, (01)00884838084926(21)US1734000893  US1736000918, (01)00884838084926(21)US1736000918  US1709000209, (01)00884838083523(21)US1709000209  US1709000241, (01)00884838083523(21)US1709000241  US1709000271, (01)00884838083523(21)US1709000271  US1709000277, (01)00884838083523(21)US1709000277  US1709000278, (01)00884838083523(21)US1709000278  US1709000281, (01)00884838083523(21)US1709000281  US1709000282, (01)00884838083523(21)US1709000282  US1709000283, (01)00884838083523(21)US1709000283  US1709000285, (01)00884838083523(21)US1709000285  US1736000946, (01)00884838083523(21)US1736000946  US1736000950, (01)00884838083523(21)US1736000950  US1738001026, (01)00884838083523(21)US1738001026  US1734000902, (01)00884838084926(21)US1734000902  US1734000906, (01)00884838084926(21)US1734000906  US1736000997, (01)00884838084926(21)US1736000997  US1738001018, (01)00884838084926(21)US1738001018  US1738001051, (01)00884838084926(21)US1738001051  US1738001052, (01)00884838084926(21)US1738001052  US1709000280, (01)00884838084926(21)US1709000280  US1709000286, (01)00884838084926(21)US1709000286  US1716000296, (01)00884838084926(21)US1716000296  US1716000300, (01)00884838084926(21)US1716000300  US1716000301, (01)00884838084926(21)US1716000301  US1716000309, (01)00884838084926(21)US1716000309  US1716000310, (01)00884838084926(21)US1716000310  US1716000313, (01)00884838084926(21)US1716000313  US1716000315, (01)00884838084926(21)US1716000315  US1708000194, (01)00884838084926(21)US1708000194  US1734000822, (01)00884838083523(21)US1734000822    US1734000782, (01)00884838084902(21)US1734000782  US1731000728, (01)00884838083523(21)US1731000728  US1731000729, (01)00884838083523(21)US1731000729  US1709000228, (01)00884838083523(21)US1709000228  US1709000231, (01)00884838083523(21)US1709000231  US1729000650, (01)00884838084902(21)US1729000650  US1708000134, (01)00884838083523(21)US1708000134  US1709000265, (01)00884838083523(21)US1709000265  US1709000266, (01)00884838083523(21)US1709000266  US1724000488, (01)00884838083530(21)US1724000488  US1724000513, (01)00884838083530(21)US1724000513  US1724000550, (01)00884838083530(21)US1724000550  US1724000555, (01)00884838083530(21)US1724000555  US1734000803, (01)00884838084933(21)US1734000803  US1734000839, (01)00884838084933(21)US1734000839  US1736000993, (01)00884838084902(21)US1736000993  US1724000486, (01)00884838083523(21)US1724000486  US1724000491, (01)00884838083523(21)US1724000491  US1724000494, (01)00884838083523(21)US1724000494  US1724000509, (01)00884838083523(21)US1724000509  US1708000178, (01)00884838083523(21)US1708000178  US1708000181, (01)00884838083523(21)US1708000181  US1708000186, (01)00884838083523(21)US1708000186  US1709000272, (01)00884838083523(21)US1709000272  US1720000351, (01)00884838083523(21)US1720000351  US1720000362, (01)00884838083523(21)US1720000362  US1720000364, (01)00884838083523(21)US1720000364    US1731000712, (01)00884838084902(21)US1731000712  US1738001102, (01)00884838084902(21)US1738001102  US1731000671, (01)00884838084902(21)US1731000671  US1736000943, (01)00884838084902(21)US1736000943  US1729000626, (01)00884838084902(21)US1729000626  US1731000714, (01)00884838084902(21)US1731000714  US1724000498, (01)00884838084902(21)US1724000498  US1729000611, (01)00884838084902(21)US1729000611  US1736000952, (01)00884838084902(21)US1736000952  US1738001023, (01)00884838084902(21)US1738001023  US1709000218, (01)00884838084902(21)US1709000218  US1736000982, (01)00884838084902(21)US1736000982  US1716000341, (01)00884838084902(21)US1716000341  US1731000705, (01)00884838084902(21)US1731000705  US1652000002, (01)00884838084926(21)US1652000002  US1652000018, (01)00884838084926(21)US1652000018  US1652000022, (01)00884838084926(21)US1652000022  US1652000046, (01)00884838084926(21)US1652000046  US1652000050, (01)00884838084926(21)US1652000050  US1652000097, (01)00884838084926(21)US1652000097  US1708000104, (01)00884838084926(21)US1708000104  US1708000106, (01)00884838084926(21)US1708000106  US1708000121, (01)00884838084926(21)US1708000121  US1708000123, (01)00884838084926(21)US1708000123  US1708000124, (01)00884838084926(21)US1708000124  US1708000125, (01)00884838084926(21)US1708000125  US1708000128, (01)00884838084926(21)US1708000128  US1708000129, (01)00884838084926(21)US1708000129  US1708000130, (01)00884838084926(21)US1708000130  US1708000131, (01)00884838084926(21)US1708000131  US1708000136, (01)00884838084926(21)US1708000136  US1708000137, (01)00884838084926(21)US1708000137  US1708000139, (01)00884838084926(21)US1708000139  US1708000141, (01)00884838084926(21)US1708000141  US1708000143, (01)00884838084926(21)US1708000143  US1708000147, (01)00884838084926(21)US1708000147  US1708000148, (01)00884838084926(21)US1708000148  US1708000151, (01)00884838084926(21)US1708000151  US1708000153, (01)00884838084926(21)US1708000153  US1708000155, (01)00884838084926(21)US1708000155  US1708000156, (01)00884838084926(21)US1708000156  US1708000157, (01)00884838084926(21)US1708000157  US1708000158, (01)00884838084926(21)US1708000158  US1708000160, (01)00884838084926(21)US1708000160  US1708000162, (01)00884838084926(21)US1708000162  US1708000165, (01)00884838084926(21)US1708000165  US1708000166, (01)00884838084926(21)US1708000166  US1708000167, (01)00884838084926(21)US1708000167  US1708000169, (01)00884838084926(21)US1708000169  US1708000172, (01)00884838084926(21)US1708000172  US1708000176, (01)00884838084926(21)US1708000176  US1708000179, (01)00884838084926(21)US1708000179  US1708000180, (01)00884838084926(21)US1708000180  US1708000182, (01)00884838084926(21)US1708000182  US1708000185, (01)00884838084926(21)US1708000185  US1708000187, (01)00884838084926(21)US1708000187  US1708000188, (01)00884838084926(21)US1708000188  US1708000189, (01)00884838084926(21)US1708000189  US1708000191, (01)00884838084926(21)US1708000191  US1708000192, (01)00884838084926(21)US1708000192  US1708000195, (01)00884838084926(21)US1708000195  US1708000196, (01)00884838084926(21)US1708000196  US1708000197, (01)00884838084926(21)US1708000197  US1708000199, (01)00884838084926(21)US1708000199  US1708000201, (01)00884838084926(21)US1708000201  US1709000204, (01)00884838084926(21)US1709000204  US1709000208, (01)00884838084926(21)US1709000208  US1709000211, (01)00884838084926(21)US1709000211  US1709000215, (01)00884838084926(21)US1709000215  US1709000227, (01)00884838084926(21)US1709000227  US1709000229, (01)00884838084926(21)US1709000229  US1709000242, (01)00884838084926(21)US1709000242  US1709000243, (01)00884838084926(21)US1709000243  US1709000246, (01)00884838084926(21)US1709000246  US1709000247, (01)00884838084926(21)US1709000247  US1709000257, (01)00884838084926(21)US1709000257  US1709000262, (01)00884838084926(21)US1709000262  US1709000267, (01)00884838084926(21)US1709000267  US1709000270, (01)00884838084926(21)US1709000270  US1716000291, (01)00884838084926(21)US1716000291  US1716000292, (01)00884838084926(21)US1716000292  US1716000298, (01)00884838084926(21)US1716000298  US1716000302, (01)00884838084926(21)US1716000302  US1716000314, (01)00884838084926(21)US1716000314  US1716000317, (01)00884838084926(21)US1716000317  US1716000318, (01)00884838084926(21)US1716000318  US1716000319, (01)00884838084926(21)US1716000319  US1716000320, (01)00884838084926(21)US1716000320  US1716000325, (01)00884838084926(21)US1716000325  US1716000328, (01)00884838084926(21)US1716000328  US1716000330, (01)00884838084926(21)US1716000330  US1716000331, (01)00884838084926(21)US1716000331  US1716000334, (01)00884838084926(21)US1716000334  US1716000336, (01)00884838084926(21)US1716000336  US1716000337, (01)00884838084926(21)US1716000337  US1720000359, (01)00884838084926(21)US1720000359  US1720000375, (01)00884838084926(21)US1720000375  US1720000379, (01)00884838084926(21)US1720000379  US1720000380, (01)00884838084926(21)US1720000380  US1720000381, (01)00884838084926(21)US1720000381  US1720000383, (01)00884838084926(21)US1720000383  US1720000385, (01)00884838084926(21)US1720000385  US1720000388, (01)00884838084926(21)US1720000388  US1720000390, (01)00884838084926(21)US1720000390  US1720000392, (01)00884838084926(21)US1720000392  US1720000393, (01)00884838084926(21)US1720000393  US1720000394, (01)00884838084926(21)US1720000394  US1720000395, (01)00884838084926(21)US1720000395  US1720000396, (01)00884838084926(21)US1720000396  US1720000397, (01)00884838084926(21)US1720000397  US1720000398, (01)00884838084926(21)US1720000398  US1720000399, (01)00884838084926(21)US1720000399  US1720000401, (01)00884838084926(21)US1720000401  US1720000402, (01)00884838084926(21)US1720000402  US1720000403, (01)00884838084926(21)US1720000403  US1720000404, (01)00884838084926(21)US1720000404  US1720000406, (01)00884838084926(21)US1720000406  US1720000408, (01)00884838084926(21)US1720000408  US1720000411, (01)00884838084926(21)US1720000411  US1720000413, (01)00884838084926(21)US1720000413  US1720000415, (01)00884838084926(21)US1720000415  US1720000418, (01)00884838084926(21)US1720000418  US1720000420, (01)00884838084926(21)US1720000420  US1720000426, (01)00884838084926(21)US1720000426  US1720000438, (01)00884838084926(21)US1720000438  US1720000440, (01)00884838084926(21)US1720000440  US1720000443, (01)00884838084926(21)US1720000443  US1720000445, (01)00884838084926(21)US1720000445  US1720000446, (01)00884838084926(21)US1720000446  US1720000449, (01)00884838084926(21)US1720000449  US1720000450, (01)00884838084926(21)US1720000450  US1720000451, (01)00884838084926(21)US1720000451  US1720000452, (01)00884838084926(21)US1720000452  US1720000454, (01)00884838084926(21)US1720000454  US1720000460, (01)00884838084926(21)US1720000460  US1720000461, (01)00884838084926(21)US1720000461  US1724000465, (01)00884838084926(21)US1724000465  US1724000467, (01)00884838084926(21)US1724000467  US1724000468, (01)00884838084926(21)US1724000468  US1724000469, (01)00884838084926(21)US1724000469  US1724000471, (01)00884838084926(21)US1724000471  US1724000472, (01)00884838084926(21)US1724000472  US1724000476, (01)00884838084926(21)US1724000476  US1724000477, (01)00884838084926(21)US1724000477  US1724000479, (01)00884838084926(21)US1724000479  US1724000483, (01)00884838084926(21)US1724000483  US1724000504, (01)00884838084926(21)US1724000504  US1724000527, (01)00884838084926(21)US1724000527  US1724000532, (01)00884838084926(21)US1724000532  US1724000536, (01)00884838084926(21)US1724000536  US1724000556, (01)00884838084926(21)US1724000556  US1724000557, (01)00884838084926(21)US1724000557  US1720000457, (01)00884838084926(21)US1720000457  US1731000673, (01)00884838084902(21)US1731000673  US1734000796, (01)00884838084902(21)US1734000796  US1729000613, (01)00884838084902(21)US1729000613  US1731000715, (01)00884838084902(21)US1731000715  US1652000043, (01)00884838084902(21)US1652000043  US1716000307, (01)00884838084902(21)US1716000307  US1709000260, (01)00884838084902(21)US1709000260  US1729000647, (01)00884838084902(21)US1729000647  US1729000564, (01)00884838084902(21)US1729000564  US1734000762, (01)00884838084902(21)US1734000762  US1716000305, (01)00884838084902(21)US1716000305  US1724000520, (01)00884838084902(21)US1724000520  US1736000947, (01)00884838084902(21)US1736000947    Model Number 989803199561  Serial Numbers/UDI (if available):  US1652000004, (01)00884838083523(21)US1652000004  US1716000327, (01)00884838083523(21)US1716000327  US1716000329, (01)00884838083523(21)US1716000329  US1716000333, (01)00884838083523(21)US1716000333  US1716000340, (01)00884838083523(21)US1716000340  US1734000892, (01)00884838083523(21)US1734000892  US1734000895, (01)00884838083523(21)US1734000895  US1738001036, (01)00884838083523(21)US1738001036  US1738001040, (01)00884838083523(21)US1738001040  US1738001043, (01)00884838083523(21)US1738001043  US1738001044, (01)00884838083523(21)US1738001044  US1738001050, (01)00884838083523(21)US1738001050  US1709000238, (01)00884838084926(21)US1709000238  US1709000252, (01)00884838084926(21)US1709000252  US1709000258, (01)00884838084926(21)US1709000258  US1729000579, (01)00884838084926(21)US1729000579  US1729000615, (01)00884838084926(21)US1729000615  US1729000617, (01)00884838084926(21)US1729000617  US1729000624, (01)00884838084926(21)US1729000624  US1729000642, (01)00884838084926(21)US1729000642  US1652000047, (01)00884838083523(21)US1652000047  US1652000001, (01)00884838083523(21)US1652000001",
      "recall_initiation_date": "20180314",
      "postal_code": "32826-3286",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}