  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Union City",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76920",
      "recalling_firm": "Mizuho OSI",
      "address_1": "30031 Ahern Ave",
      "address_2": "N/A",
      "postal_code": "94587-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United States, Canada, Japan, Australia, New Zealand, Great Britain, Sweden, Spain, Germany, Switzerland, Kenya, Saudi Arabia, Hong Kong, South Korea.",
      "recall_number": "Z-1867-2017",
      "product_description": "216 Wilson Frame Pads, Model # 5319-37;    Used with Radiolucent Wilson Frames, model numbers 5319G and 5321G.    General and Plastic Surgery: These pads are used as a component of the Wilson Frames, which support the torso of a prone patient during spine surgery. The pads are supplied with the frames and as replacements.",
      "product_quantity": "216 sets",
      "reason_for_recall": "Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.",
      "recall_initiation_date": "20170328",
      "center_classification_date": "20170419",
      "termination_date": "20170815",
      "report_date": "20170426",
      "code_info": "All Lots:  Manufacturing dates: December 2016 through March 2017;  GTIN:00842430100192"
    }
  ]
}