  "results": [
    {
      "status": "Terminated",
      "city": "Bozeman",
      "state": "MT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65544",
      "recalling_firm": "Mesa Laboratories, Inc., Bozeman Manufacturing Facility",
      "address_1": "10 Evergreen Dr Ste E",
      "address_2": "N/A",
      "postal_code": "59715-2457",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Australia, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Egypt, Greece, India,Israel, Japan, Jordan, Malaysia, Mexico, Milanese, Netherlands, Peru, Portugal, Pozan, Romania, Spain, Sweden, Taiwan, Thailand, United Kingdom, and Uruguay.",
      "recall_number": "Z-1867-2013",
      "product_description": "Mesa EZtest BIOLOGICAL INDICATOR MONITORING SYSTEM.     Intended for use in determining the efficacy of steam sterilization processes.",
      "product_quantity": "865/25 ct boxes, 414/100 ct boxes, 307/500 ct boxes",
      "reason_for_recall": "Mesa EZTest , Self-contained steam biological indicator, are recalled because Mesa recently determined that the read out time is not in specification with the label claims.",
      "recall_initiation_date": "20130604",
      "center_classification_date": "20130801",
      "termination_date": "20131122",
      "report_date": "20130807",
      "code_info": "Lot No S-444 Manufacture Date 2013 January 21 Expiration Date 2015 January 21;    Lot No S-445 Manufacture Date 2013 February 15 Expiration Date 2015 February 15 ;    Lot No S-447 Manufacture Date 2013 March 18 Expiration Date 2015 March 18.",
      "more_code_info": ""
    }
  ]
}