{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marietta",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82903",
      "recalling_firm": "Thermo Fisher Scientific (Asheville) LLC",
      "address_1": "401 Mill Creek Rd",
      "address_2": "N/A",
      "postal_code": "45750-4304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "CA, GA, KY, PA  Foreign: Canada, China, Germany, Japan, New Zealand, Australia",
      "recall_number": "Z-1865-2019",
      "product_description": "Fisher Scientific Isotemp -General Purpose Water Bath:   lsotemp 5L GP Bath  Model Number: FSGPD05",
      "product_quantity": "10 units",
      "reason_for_recall": "Product may have been assembled incorrectly,the  assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.",
      "recall_initiation_date": "20190402",
      "center_classification_date": "20190621",
      "termination_date": "20200723",
      "report_date": "20190703",
      "code_info": "Serial Numbers:  30024679S  300246796  300246797  300246798  300246799  300247393  300247394  30024739S  300247396  3002477S8"
    }
  ]
}