  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Batesville",
      "address_1": "1069 State Route 46 East",
      "reason_for_recall": "During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail.",
      "address_2": "",
      "product_quantity": "7843 devices",
      "code_info": "AdvantaTM 2 beds (P1190) produced between August 10, 2009 and October 6, 2011; bed S/N between K222AT9180 and M279AT4403",
      "center_classification_date": "20120622",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Canada, France, and Singapore.",
      "state": "IN",
      "product_description": "Advanta 2 Bed.    The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital.",
      "report_date": "20120704",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hill-Rom, Inc.",
      "recall_number": "Z-1864-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62331",
      "termination_date": "20140606",
      "more_code_info": "",
      "recall_initiation_date": "20120618",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}