  "results": [
    {
      "status": "Terminated",
      "city": "Lawrence",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68354",
      "recalling_firm": "NxStage Medical, Inc.",
      "address_1": "350 Merrimack Street",
      "address_2": "N/A",
      "postal_code": "01843-1748",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to France, Netherlands, Spain, Sweden, and the UK.",
      "recall_number": "Z-1860-2014",
      "product_description": "NxStage Dialysate Sack (SAK) with Concentrate  for Use with PureFlow SL  REF: SAK-307  Lactate 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1,   Chloride 105, Glucose (mg/dL) 100  Batch Size (liters) 50    The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.",
      "product_quantity": "8683 cases",
      "reason_for_recall": "SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification",
      "recall_initiation_date": "20140514",
      "center_classification_date": "20140624",
      "termination_date": "20170313",
      "report_date": "20140702",
      "code_info": "Lot code:  40179126  40179125  40179108  40179036  40179035  31279119  3057923  3057907  3047927"
    }
  ]
}