  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76507",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide in the   ARGENTINA                  AUSTRALIA                  CANADA  CHILE                      CHINA                      COSTA RICA                 ECUADOR                    GSCC  INDIA                      ITALY                      JAPAN                      KOREA, THE REP OF (SOUTH)  MALAYSIA  MEXICO                     PANAMA                     Singapore  SWITZERLAND  THAILAND  UK",
      "recall_number": "Z-1858-2017",
      "product_description": "various polyethylene implants   BIO PLUG 10MM      Product Usage:    Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant.",
      "product_quantity": "28,253 in total",
      "reason_for_recall": "Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.",
      "recall_initiation_date": "20170217",
      "center_classification_date": "20170418",
      "termination_date": "20180129",
      "report_date": "20170426",
      "code_info": "Item 163005  Lot 309030"
    }
  ]
}