{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82978",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, ALBANIA GHANA PORTUGAL ARGENTINA HONG KONG ROMANIA  AUSTRIA HUNGARY RUSSIA AUSTRALIA INDIA SAUDI ARABIA  AZERBAIJAN IRAN SINGAPORE BAHAMAS IRELAND SLOVAKIA  BANGLADESH IRAQ SLOVENIA BELGIUM ISRAEL SOUTH AFRICA  BOLIVIA ITALY SOUTH KOREA BRAZIL JAPAN SPAIN BULGARIA JORDAN SWEDEN CAMEROON LATVIA SWITZERLAND CANADA LEBANON TAIWAN  CHILE LUXEMBOURG TANZANIA CHINA MALAYSIA THAILAND  COLOMBIA MEXICO TRINIDAD and TOBAGO COSTA RICA MOLDAVA TURKEY CROATIA MOZAMBIQUE UNITED ARAB EMIRATES  CUBA MYANMAR UNITED KINGDOM CZECH REPUBLIC NETHERLANDS VIETNAM DENMARK NAMIBIA VENEZUELA DOMINICAN REPUBLIC NIGERIA YEMEN ECUADOR NORWAY ZIMBABWE EGYPT NEW ZEALAND ESTONIA OMAN ESWATINI PAKISTAN FINLAND PANAMA FRANCE PARAGUAY GERMANY POLAND",
      "recall_number": "Z-1857-2019",
      "product_description": "Maquet VOLISTA StandOP Surgical Lights-  Product Code: ARDVST229000A  Model: VST60SF AIM",
      "product_quantity": "2 units",
      "reason_for_recall": "Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk  of particles falling into the surgical field during surgery that may result in patient infection",
      "recall_initiation_date": "20190528",
      "center_classification_date": "20190621",
      "termination_date": "20230710",
      "report_date": "20190703",
      "code_info": "Serial Numbers:   500000 500001"
    }
  ]
}