  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Yorba Linda",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85159",
      "recalling_firm": "Nobel Biocare Usa Llc",
      "address_1": "22715 Savi Ranch Pkwy",
      "address_2": "N/A",
      "postal_code": "92887-4609",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AZ, AR, CA, CO, FL, GA, IL, IN, ID, KS, LA, MI, MD, ME, MA, MO, NY, NC, NJ, NE, NM, OH, OR, OK, PA, TX, TN, VA, WY, WA and WI. Countries of CA, CH, DE, ES, FR, GB, HU, IT, BR, CO, MX, CN, HK, IN, JP and TW.",
      "recall_number": "Z-1856-2020",
      "product_description": "Universal Base CC RP 1.5 mm; The Universal Base Abutment in combination with Nobel Biocare end osseous implants are indicated for single-unit if screw retained and for multiple-unit if cement retained crowns are used; catalog no. 38214 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.",
      "product_quantity": "238 pieces",
      "reason_for_recall": "Packages of abutments (universal base) may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.",
      "recall_initiation_date": "20190809",
      "center_classification_date": "20200504",
      "termination_date": "20211201",
      "report_date": "20200513",
      "code_info": "Catalog no. 38214, Lot no. 13076822, UDI: (01)07332747096538(10)13076822",
      "more_code_info": ""
    }
  ]
}