  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82978",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, ALBANIA GHANA PORTUGAL ARGENTINA HONG KONG ROMANIA  AUSTRIA HUNGARY RUSSIA AUSTRALIA INDIA SAUDI ARABIA  AZERBAIJAN IRAN SINGAPORE BAHAMAS IRELAND SLOVAKIA  BANGLADESH IRAQ SLOVENIA BELGIUM ISRAEL SOUTH AFRICA  BOLIVIA ITALY SOUTH KOREA BRAZIL JAPAN SPAIN BULGARIA JORDAN SWEDEN CAMEROON LATVIA SWITZERLAND CANADA LEBANON TAIWAN  CHILE LUXEMBOURG TANZANIA CHINA MALAYSIA THAILAND  COLOMBIA MEXICO TRINIDAD and TOBAGO COSTA RICA MOLDAVA TURKEY CROATIA MOZAMBIQUE UNITED ARAB EMIRATES  CUBA MYANMAR UNITED KINGDOM CZECH REPUBLIC NETHERLANDS VIETNAM DENMARK NAMIBIA VENEZUELA DOMINICAN REPUBLIC NIGERIA YEMEN ECUADOR NORWAY ZIMBABWE EGYPT NEW ZEALAND ESTONIA OMAN ESWATINI PAKISTAN FINLAND PANAMA FRANCE PARAGUAY GERMANY POLAND",
      "recall_number": "Z-1854-2019",
      "product_description": "Maquet VOLISTA StandOP Surgical Lights-  Product Code: ARD568811911  Model: VLT600SF AIM STP",
      "product_quantity": "65 units",
      "reason_for_recall": "Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk  of particles falling into the surgical field during surgery that may result in patient infection",
      "recall_initiation_date": "20190528",
      "center_classification_date": "20190621",
      "termination_date": "20230710",
      "report_date": "20190703",
      "code_info": "Serial Numbers:   31000 31001 31002 31003  31004 31006 31008 31009  31010 31012 31017 31018  31019 31020 31021 31022  31023 31024 31025 31026  35000 35001 35011 35012  35013 35014 35015 35017  35018 35019 35020 35021  35022 35023 35024 35025  35026 35027 35028 35030  35031 35032 35033 38000  38001 38002 38003 38009  38011 38012 38013 38014  38015 38016 38018 38019  38020 38021 38022 38023  38025 38026 38027"
    }
  ]
}