  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92373",
      "recalling_firm": "TELEFLEX LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including Puerto Rico.",
      "recall_number": "Z-1853-2023",
      "product_description": "Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590",
      "product_quantity": "13710 units",
      "reason_for_recall": "Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.",
      "recall_initiation_date": "20230525",
      "center_classification_date": "20230629",
      "report_date": "20230705",
      "code_info": "UDI/DI 4026704547654, Batch Numbers: 18FG18, 18FG20, 18GG17, 18GT32, 18HG12, 18HG27, 18HG38, 18IG15, 18IG30, 18KG03, 18KG07, 18KG35, 18LG18, 19AG24, 19AG27, 19BG04, 19DT38, 19DT45, 19ET32, 19ET72, 19FT14, 19GT20, 19HT40, 19IT39, 19IT58, 19KT17, 19KT21, 19LT50, 20BG16, 20BT05, 20BT47, 20CT36, 20CT44, KME20H0238, KME20H0239, KME20H0591, KME20J2420, KME21E1114, KME22B1404, KME22B1405, KME22B1755, KME22C0915, KME22C1943, KME22C2372, KME22D1717, KME22D1718, KME22G1443, KME23A2760, KME23C0137",
      "more_code_info": ""
    }
  ]
}