  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82822",
      "recalling_firm": "Tecan US, Inc.",
      "address_1": "9401 Globe Center Dr Ste 140",
      "address_2": "N/A",
      "postal_code": "27560-6211",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US Nationwide distribution in the states CA, CO, CT, KS, NC, NJ, OR, RI, TN, TX, and VA. Country of Canada.",
      "recall_number": "Z-1851-2019",
      "product_description": "Freedom EVO 200 - Product Usage: It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.",
      "product_quantity": "40",
      "reason_for_recall": "Gas springs not always replaced within the interval specified.",
      "recall_initiation_date": "20190412",
      "center_classification_date": "20190621",
      "termination_date": "20220118",
      "report_date": "20190703",
      "code_info": "Freedom EVO 200 BASE UNIT 10641200 and 10641541  Freedom EVO 200 MCA96 BASE UNIT  30020020  Freedom EVO 200 MCA384 BASE UNIT 30032020  Freedom EVO 200-2 LiHa BASE UNIT  10641202    Serial Numbers  222, 702, 714, 506000013, 511000005, 603000004, 702000004, 703000002, 703000005, 703000006, 707001360, 710000367, 801002513, 801003121, 802003688, 805002719, 809002562, 810000017, 812002002, 906005647, 908002944, 910006632, 1004001243, 1006000274, 1006009336, 1009002730, 1011003146, 1011004081, 1011004086, 1011004089, 1011004095, 1104000151, 1108006349, 1108006724, 1108006777, 1109000886, 1109001036, 1109006938, 1111008689, and 1203005259."
    }
  ]
}