  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Charlottesville",
      "address_1": "501 Locust Ave",
      "reason_for_recall": "After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the  system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).",
      "address_2": "",
      "product_quantity": "55",
      "code_info": "JAQ",
      "center_classification_date": "20150626",
      "distribution_pattern": "Worldwide Distribution.",
      "state": "VA",
      "product_description": "The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Varian Medical Systems Inc",
      "recall_number": "Z-1849-2015",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "71451",
      "termination_date": "20160912",
      "more_code_info": "",
      "recall_initiation_date": "20150518",
      "postal_code": "22902-4869",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}