{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87984",
      "recalling_firm": "Brainlab AG",
      "address_1": "Olof-palme-str. 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, AR, CA, FL, IN, NJ, NY, NC, OR, PA, and TX.  There was no government/military distribution.",
      "recall_number": "Z-1846-2021",
      "product_description": "ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.",
      "product_quantity": "22 systems in the U.S.",
      "reason_for_recall": "Display of potential patient movement might be delayed to the user for high dose treatments.",
      "recall_initiation_date": "20210514",
      "center_classification_date": "20210610",
      "termination_date": "20230320",
      "report_date": "20210616",
      "code_info": "Version 1.02, GTIN 04056481142315"
    }
  ]
}