  "results": [
    {
      "status": "Terminated",
      "city": "Rancho Santa Margarita",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85057",
      "recalling_firm": "Applied Medical Resources Corp",
      "address_1": "22872 Avenida Empresa",
      "address_2": "N/A",
      "postal_code": "92688-2650",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - United States: Nationwide    OUS: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, South Africa, Spain, Taiwan, United Kingdom",
      "recall_number": "Z-1846-2020",
      "product_description": "Python Embolectomy Catheters, Sterile EO, CE, Ref numbers/UDI: A4E01/(01)00607915113407;  A4E02/(01)00607915113414;  A4E03/(01)00607915113421;  A4E04/(01)00607915113438;  A4E05/(01)00607915113445;  A4E06/(01)00607915113452;  A4E08/(01)00607915113476;  A4E09/(01)00607915113483",
      "product_quantity": "28,799 catheters in total",
      "reason_for_recall": "The firm has received reports of tip separation during usage of  the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.",
      "recall_initiation_date": "20191024",
      "center_classification_date": "20200507",
      "termination_date": "20230601",
      "report_date": "20200513",
      "code_info": "Model A4E01, PYTHON 5MM-40CM, lot numbers 1259463, 1268371, 1274343, 1281314, 1288397, 1291497, 1302766, 1305919, 1312958, 1317578, 1319209, 1324383, 1326383, 1333050, 1338309, 1339544.  Model A4E02, PYTHON 5MM-80CM, lot numbers 1257416, 1268372, 1281315, 1283401, 1293028, 1294897, 1303013, 1305920, 1317579, 1319807, 1324384, 1331435, 1332003, 1338732, 1339490  Model A4E03, PYTHON 9MM-40CM, lot numbers 1256874,1263093,1269005,1276315,1281787,1287079,1298430,1303017, 1308139, 1319211, 1330447, 1331896, 1333407, 1337644, 1341362 (note: lot numbers edited for correctness 4/23/2021)   Model A4E04, PYTHON 9MM-80CM, lot numbers 1255539, 1262756, 1269004, 1276316, 1281786, 1288847, 1293169, 1299257, 1303808, 1312595, 1319212, 1321387, 1325381, 1331884, 1332592, 1337686, 1339491  Model A4E05, PYTHON 11MM-40CM, lot numbers 1263092, 1268170, 1273184, 1278603, 1283650, 1288347, 1292823, 1296020, 1299013, 1305922, 1310451, 1318835, 1320036, 1323535, 1325777, 1330446, 1331328, 1337641, 1339492  Model A4E06, PYTHON 11MM-80CM, lot numbers 1261855, 1268451, 1276317, 1284924, 1290454, 1298079, 1305921, 1310623, 1320183, 1323280, 1330449, 1332579, 1337652, 1339493  Model A4E08, PYTHON 13MM-80CM, lot numbers 1252636, 1259464, 1263484, 1265630, 1266989, 1267854, 1268688, 1276314, 1281784, 1283399, 1287076, 1290936, 1294898, 1303211, 1306288, 1308127, 1312957, 1315919, 1317577, 1319637, 1320236, 1324382, 1332433, 1335657, 1339494  Model A4E09, PYTHON 14MM-80CM, lot numbers 1260019, 1273374, 1281783, 1283770, 1295466, 1299012, 1305368, 1317203, 1317876, 1319210, 1324390, 1330448, 1332593, 1337653, 1339495",
      "more_code_info": ""
    }
  ]
}