  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74128",
      "recalling_firm": "Skytron, Div. The KMW Group, Inc",
      "address_1": "5085 Corporate Exchange Blvd SE",
      "address_2": "N/A",
      "postal_code": "49512-5515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.",
      "recall_number": "Z-1844-2016",
      "product_description": "EZ Slide Battery/line powered hydraulic table with removable leg section and 21\u001d top slide   General Purpose Surgical Table    Skytron EZ Slide Operating Table  General Purpose Surgical Table",
      "product_quantity": "458",
      "reason_for_recall": "The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the table\u0019s elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.",
      "recall_initiation_date": "20160406",
      "center_classification_date": "20160531",
      "termination_date": "20170720",
      "report_date": "20160608",
      "code_info": "Model number 3502.  All units.  Units entered commercial distribution between April 30, 2014 and July 24, 2014"
    }
  ]
}