{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bothell",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84902",
      "recalling_firm": "Philips Ultrasound Inc",
      "address_1": "22100 Bothell Everett Hwy",
      "address_2": "N/A",
      "postal_code": "98021-8431",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution.",
      "recall_number": "Z-1841-2020",
      "product_description": "Transducer S7-3t    UDI (01)00884838061439  REF 989605406772    Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies",
      "product_quantity": "N/A",
      "reason_for_recall": "The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury",
      "recall_initiation_date": "20190830",
      "center_classification_date": "20200501",
      "termination_date": "20241127",
      "report_date": "20200513",
      "code_info": "All lots.",
      "more_code_info": ""
    }
  ]
}