  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68379",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-1840-2014",
      "product_description": "Siemens LANTIS System ; LANTIS Commander, 10 Users ; LANTIS 6.1 Commander ; LANTIS 8.3 Commander    The intended use of the linear accelerator is to deliver x-ray radiation for therapeutic treatment of cancer.",
      "product_quantity": "108",
      "reason_for_recall": "There is a potential risk when using LANTIS OIS System client software with operating systems for which it has not been validated and released, resulting in incorrectly stored data, such as treatment records, and LANTIS database corruption.",
      "recall_initiation_date": "20140506",
      "center_classification_date": "20140620",
      "termination_date": "20141215",
      "report_date": "20140702",
      "code_info": "Material nos.: 4503178, 5493072, and 8148178 with serial nos.:  93-0695  93-0693  93-0938  93-1010  93-0730  93-0454  93-0759  93-0848  93-0884  93-1158  93-0584  93-0037  93-1053  93-0726  93-0588  93-0614  93-0566  93-1038  93-0648  93-0803  93-0813  93-1387  93-0278  93-0856  93-0314  93-0657  93-0222  93-1438  93-0662  93-0779  93-0732  93-0841  93-1045  93-0487  93-0564  93-0577  93-0792  93-0231  93-0096  93-0846  93-0574  93-0668  93-0634  93-0887  93-1145  93-0818  93-0102  93-0764  93-0501  93-0490  93-0261  93-0570  93-0739  93-1133  93-0159  93-0872  93-1531  93-0660  93-0798  93-0528  93-0326  93-0393  93-0738  93-0545  93-1296  93-0623  93-0370  93-1381  93-0951  93-0667  93-1135  93-0141  93-0786  93-0483  93-0385  93-0694  93-0883  93-1314  93-1320  93-0799  93-0456  93-0416  93-0838  93-0639  93-0659  93-0760  93-0608  93-0766  93-0912  93-0063  93-0262  93-0878  93-0861  93-0452  93-0533  93-0272  93-0724  93-0596  93-0831  93-0398  93-0805  93-0870  93-0874  93-0789  93-0686  93-0475  93-0748  93-1185"
    }
  ]
}