{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74188",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution. US, Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.",
      "recall_number": "Z-1837-2016",
      "product_description": "SEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05, 470380-06; Part number:470380.  The da Vinci Xi Surgical System cannula seal is used with the cannula accessory and provides sterile disposable cap for each cannula. When attached to the cannula, the seal maintains insufflation when the cannula is empty and when the instrument is inserted during a surgical procedure.",
      "product_quantity": "46,191 boxes total - both sizes",
      "reason_for_recall": "Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula  lumen when insufflation is connected during a surgical procedure.",
      "recall_initiation_date": "20160509",
      "center_classification_date": "20160527",
      "termination_date": "20161101",
      "report_date": "20160608",
      "code_info": "All lots"
    }
  ]
}