  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82950",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution, worldwide foreign distribution.",
      "recall_number": "Z-1836-2019",
      "product_description": "SIGNIA TRI-STAPLE 2.0 CURVED TIP 60MM MEDIUM/ THIC  , SIG60CTAMT",
      "product_quantity": "3,113,280",
      "reason_for_recall": "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.",
      "recall_initiation_date": "20190517",
      "center_classification_date": "20190615",
      "report_date": "20190626",
      "code_info": "N6E0295UX  N6H0576UX  N6J1092UX  N6K0230UX  N6K0938UX  N6L0422URX  N6L0777UX  N6M0607UX  N7A0415UX  N7C0112X  N7D1013X  N7E1089X  N7F0275X  N7F1126X  N7G0444X  N7H0047X  N7H1067X  N7J0175X  N7L0825X  N8A0249X  N8A0655X  N8B0455X  N8B0945X  N8C0757X  N8C1213X  N8D0593X  N8D0907X  N8E0015X  N8G0648X  N8J0764X  N8K0823KY  N8K0840Y  N8L0166Y  N8L0436Y  N8M0845Y  N9A0204Y  N9B0455Y"
    }
  ]
}