  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82950",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution, worldwide foreign distribution.",
      "recall_number": "Z-1834-2019",
      "product_description": "SIGNIA TRI-STAPLE 2.0 BLACK 60MM EXTRA THICK RELOA, SIG60AXT",
      "product_quantity": "3,113,280",
      "reason_for_recall": "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.",
      "recall_initiation_date": "20190517",
      "center_classification_date": "20190615",
      "report_date": "20190626",
      "code_info": "N6C0533UX  N6C0582UX  N6C0657UX  N6C0681UX  N6E0115UX  N6E0116UX  N6E0976UX  N6E0977UX  N6E0978UX  N6E0979UX  N6F0045UX  N6F0046UX  N6F0047UX  N6F0048UX  N6F0049UX  N6F0050UX  N6F0295UX  N6F0503UX  N6F0504UX  N6F0505UX  N6F0506UX  N6F0507UX  N6F0508UX  N6F0509UX  N6F0748UX  N6F0758UX  N6F0759UX  N6F0760UX  N6F0786UX  N6F0787UX  N6F0788UX  N6G0068UX  N6G0270UX  N6G0271UX  N6G0287UX  N6G0288UX  N6G0486UX  N6G0499UX  N6G0784UX  N6G0785UX  N6G0786UX  N6H0105UX  N6H0106UX  N6H0107UX  N6H0155UX  N6H0156UX  N6H0157UX  N6H0173UX  N6H0174UX  N6H0298UX  N6H0299UX  N6H0302UX  N6H0522UX  N6H0693UX  N6H0734UX  N6H0909UX  N6H0946UX  N6J0040UX  N6J0092UX  N6K0302UX  N6K0970UX  N6K1067UX  N6L0152UX  N6M0328UX  N7A0170UX  N7C0475X  N7C1091X  N7C1092X  N7D0104X  N8A0087X  N8D0035X  N8D0225X"
    }
  ]
}