  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65440",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution, including Nationwide (US) and the countries of China, Italy, Japan, Canada, Turkey, United Kingdom, Singapore, Israel, and the Netherlands.",
      "recall_number": "Z-1834-2013",
      "product_description": "AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is for in vitro diagnostic use only.",
      "product_quantity": "76 units",
      "reason_for_recall": "The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit.  This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.",
      "recall_initiation_date": "20130603",
      "center_classification_date": "20130731",
      "termination_date": "20140823",
      "report_date": "20130807",
      "code_info": "Model Numbers:  AU5811-01, AU5821-01, AU5831-01, AU5841-01,  AU5811-02, AU5821-02, AU5831-02, AU5841-02,  AU5811-03, AU5821-03, AU5831-03, AU5841-03,  AU5811-04, AU5821-04, AU5831-04, AU5841-04,  AU5811-06, AU5821-06, AU5831-06, AU5841-06    Part Numbers:  A94906, A94911, A94916, A91921  A94907, A94912, A94917, A94922  A94908, A94913, A94918, A94923  A94909, A94914, A91949, A94924  A94910, A94915, A94920, A94925    Serial Numbers: All",
      "more_code_info": ""
    }
  ]
}