      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82950",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution, worldwide foreign distribution.",
      "recall_number": "Z-1832-2019",
      "product_description": "SIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD  , SIG30AMT",
      "product_quantity": "3,113,280",
      "reason_for_recall": "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.",
      "recall_initiation_date": "20190517",
      "center_classification_date": "20190615",
      "report_date": "20190626",
      "code_info": "N6E0033UX  N6H0768UX  N6H0772UX  N6K0521UX  N6K0522UX  N6K0667UX  N6K0668UX  N6L0687UX  N6M0399UX  N7A0018UX  N7A0213UX  N7A0401UX  N7A0837UX  N7B0787UX  N7C0238X  N7C0470X  N7E0496X  N7E1088X  N7F0670X  N7H0962X  N7J0370X  N8A0188X  N8B0057X  N8B0483X  N8B0940X  N8D0365X  N8D0649X  N8E0092X  N8E0213X  N8E0516X  N8E1178X  N8F0483X  N8F0819X  N8F0974X  N8J0440X  N8L0589Y  N8M0529Y  N9A0050Y  N9A0427Y"
    }
  ]
}