  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65568",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Distribution including the states of AR, CA, FL, NJ, NY, OK, and TX",
      "recall_number": "Z-1832-2013",
      "product_description": "Exactech Articulated Driver     A nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery.",
      "product_quantity": "57",
      "reason_for_recall": "When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.",
      "recall_initiation_date": "20130530",
      "center_classification_date": "20130731",
      "termination_date": "20130919",
      "report_date": "20130807",
      "code_info": "Lot #: 52667001  Catalog #: 181-00-04",
      "more_code_info": ""
    }
  ]
}