  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grove City",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65576",
      "recalling_firm": "Tosoh Smd Inc",
      "address_1": "3600 Gantz Rd",
      "address_2": "N/A",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.  Internationally to Canada, Japan, Mexico, Chile, and Uruguay.",
      "recall_number": "Z-1830-2013",
      "product_description": "AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag    AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.",
      "product_quantity": "11,309 bags",
      "reason_for_recall": "Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.",
      "recall_initiation_date": "20130515",
      "center_classification_date": "20130730",
      "termination_date": "20131127",
      "report_date": "20130807",
      "code_info": "The following lots numbers are affected by this recall:     060811, 070111, B11, B11-12/23/2011, B12-02/07/2012, B12-04/30/2012, B12-06/01/2012, B12-08/06/2012, B12-09/05/2012, B12-10/26/2012, and B12-11/27/2012.",
      "more_code_info": ""
    }
  ]
}