  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Charlottesville",
      "address_1": "501 Locust Ave",
      "reason_for_recall": "Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization  GM11010080 2012-09-06.     The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat",
      "address_2": "",
      "product_quantity": "46",
      "code_info": "JAQ",
      "center_classification_date": "20150623",
      "distribution_pattern": "Worldwide Distribution.",
      "state": "VA",
      "product_description": "Surface Applicator Set with Leipzig-style Cone.    Brachytherapy applicator set.",
      "report_date": "20150701",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Varian Medical Systems Inc",
      "recall_number": "Z-1829-2015",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "71450",
      "termination_date": "20160126",
      "more_code_info": "",
      "recall_initiation_date": "20150518",
      "postal_code": "22902-4869",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}