{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Englewood",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65473",
      "recalling_firm": "Steritec Products Mfc. Co., Inc.",
      "address_1": "74 Inverness Drive East",
      "address_2": "N/A",
      "postal_code": "80112-5114",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (Nationwide) and countries of: Argentina, Australia, Canada, British Virgin Island, Czech Republic, Italy, and Turkey.",
      "recall_number": "Z-1829-2013",
      "product_description": "Cardinal BI Stat Biological Test Pack with Instant Readout Integrator, SteriTec Product Code PL 160/A, Distributor Product Code T40311PE    SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132¿C (270¿F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome.",
      "product_quantity": "360 units",
      "reason_for_recall": "Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.",
      "recall_initiation_date": "20130610",
      "center_classification_date": "20130729",
      "termination_date": "20131219",
      "report_date": "20130807",
      "code_info": "Affected Biological Indicator Lot No.:130401",
      "more_code_info": ""
    }
  ]
}