      "status": "Terminated",
      "city": "Englewood",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65473",
      "recalling_firm": "Steritec Products Mfc. Co., Inc.",
      "address_1": "74 Inverness Drive East",
      "address_2": "N/A",
      "postal_code": "80112-5114",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (Nationwide) and countries of: Argentina, Australia, Canada, British Virgin Island, Czech Republic, Italy, and Turkey.",
      "recall_number": "Z-1828-2013",
      "product_description": "Medline Med-Checks BI Test packs with Instant Readout Integrator, SteriTec Product Code PL 200, Distributor Product Code MDS200500IR    SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132¿C (270¿F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome",
      "product_quantity": "3535 units",
      "reason_for_recall": "Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.",
      "recall_initiation_date": "20130610",
      "center_classification_date": "20130729",
      "termination_date": "20131219",
      "report_date": "20130807",
      "code_info": "Affected Biological Indicator Lot No.:130401",
      "more_code_info": ""
    }
  ]
}