      "country": "United States",
      "city": "South Jordan",
      "address_1": "1600 West Merit Parkway",
      "reason_for_recall": "Some of these introducer sheaths were packaged with a 0.035 inch dilator, rather than the labeled 0.038 inch. This affects one lot, H767887, of catalog number PR0-6F-11-038.",
      "address_2": "",
      "product_quantity": "1,445",
      "code_info": "Catalo Number PR0-6F-11-038, Lot Number H767887",
      "center_classification_date": "20150622",
      "distribution_pattern": "Worldwide Distribution - US including TN, NE, MA, SC, WI, UT and Internationally to Belgium, Switzerland, United Kingdom, and Germany.",
      "state": "UT",
      "product_description": "Merit Medical Prelude Pro Sheath Introducter, 6F (2.0mm) 11cm, REF PRO-6F-11-038, Sterile EO  The Merit Prelude Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.",
      "report_date": "20150701",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Merit Medical Systems, Inc.",
      "recall_number": "Z-1826-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71519",
      "termination_date": "20151026",
      "more_code_info": "",
      "recall_initiation_date": "20150604",
      "postal_code": "84095",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}