{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98588",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.",
      "recall_number": "Z-1825-2026",
      "product_description": "Philips Spectral CT on Rails. Model Number: 728334.",
      "product_quantity": "3 systems",
      "reason_for_recall": "Philips has identified three software issues:  1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go.   2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear.   3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.",
      "recall_initiation_date": "20260307",
      "center_classification_date": "20260416",
      "report_date": "20260422",
      "code_info": "Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007."
    }
  ]
}