{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orange",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85284",
      "recalling_firm": "Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical",
      "address_1": "1717 W Collins Ave",
      "address_2": "N/A",
      "postal_code": "92867-5422",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of NV, TN, NY, WA, WI, GA, PA, TX, IN, FL, IL, IA, MD, CA, LA, , MI, AZ, NC, KY, SC and the country of Netherlands.",
      "recall_number": "Z-1824-2020",
      "product_description": "Nexus Universal with OptiBond Universal, Universal Adhesive Resin Cement, Trial Kit, Ref #  36990 - Product Usage: The intended use of this device is to be used for the cementation of indirect restorations including veneers, inlays, onlays, bridges, posts, and the cementation of crown restorations to implant abutments.",
      "product_quantity": "111 kits",
      "reason_for_recall": "An issue within  production process used to package trial kits let to the inclusion of the incorrect bonding agent for the kit. Use of the incorrect bonding agent may lead to increased risk of bond failures.",
      "recall_initiation_date": "20200225",
      "center_classification_date": "20200429",
      "termination_date": "20210519",
      "report_date": "20200506",
      "code_info": "Lot Numbers: 7402423, 7290257, 7451039"
    }
  ]
}