  "results": [
    {
      "status": "Terminated",
      "city": "Uijeongbu-Si Gyeonggi-Do",
      "state": "N/A",
      "country": "Korea (the Republic of)",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84989",
      "recalling_firm": "U&I CORP.",
      "address_1": "20 Sandan-Ro 76 Beon-Gil",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of CA.",
      "recall_number": "Z-1823-2020",
      "product_description": "Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.",
      "product_quantity": "42",
      "reason_for_recall": "The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.",
      "recall_initiation_date": "20190130",
      "center_classification_date": "20200429",
      "termination_date": "20201120",
      "report_date": "20200506",
      "code_info": "Lots: A18K220, A18L065, A18L013, A18L014, A18L015, A18L018, A18L019, A18L020, A18L021"
    }
  ]
}