  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Miamisburg",
      "address_1": "452 Alexandersville Rd",
      "reason_for_recall": "The Continuous Radius Rod should measure 50mm in length but actually measures 45mm in length.",
      "address_2": "",
      "product_quantity": "75 units",
      "code_info": "Model #: N60001068; Lot #: 2943-01; Serial #\u0019s: 46118, 46581, 47296, 47588, 47653, 48164, 48201, 48412, 49156, 49585, 49586, 49649, G22, G45, G47, G77, G83, G87X, G90X, R15, XP03, XP21, XP27, XP28, XP29 & XP32.",
      "center_classification_date": "20150622",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of : FL, ID, IL, KS, LA, MI, MO, NJ, OH, PA and TX., and to the countries of Austria, Germany and Italy.",
      "state": "OH",
      "product_description": "Fortex\" Pedicle Screw System Continuous Radius Rod 5.5 - 50mm, X-Spine Systems, Inc., 452 Alexandersville Road Miamisburg, OH 45342 Phone: 937-847-8400",
      "report_date": "20150701",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "X Spine Systems Inc",
      "recall_number": "Z-1823-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71357",
      "termination_date": "20151001",
      "more_code_info": "",
      "recall_initiation_date": "20150407",
      "postal_code": "45342-3658",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}