  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90782",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV  Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.",
      "recall_number": "Z-1822-2022",
      "product_description": "Humidifier bottle and syringe filter kit (6 count of each) for the BT37  Mark I Benchtop Incubator (AY102295).    The Mark I model is no longer produced, but the legacy humidifier  bottle component design can also be used with the BT37 Mark II  Benchtop Incubator.",
      "product_quantity": "5,182 kits",
      "reason_for_recall": "There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.",
      "recall_initiation_date": "20220812",
      "center_classification_date": "20220928",
      "report_date": "20221005",
      "code_info": "Lot Numbers  05593V291121 18-1192 1812-68 18-1268 G000575  G003811  G003893  G004517  G004874  G005109  G005476  G005723  E160415  Expiration date range: 27 Jun 2022 to 07 Dec 2024"
    }
  ]
}